Cystatin C assayed from stored serum samples from NHANES 1999-2000.
All participants aged 60+ with available specimen and a 25% random sample of participants aged 12-59 years. This younger group is supplemented with all individuals with high serum creatinine 1.2 mg/dl in men and >1.0 mg/dl in women) after calibration.
All assays were conducted using the Dade Behring N Latex Cystatin C assay which is an automated particle-enhanced nephelometric assay (PENIA) run on the Dade Behring Nephelometer II (BNII) (1).
All data will be made publicly available.
This assay has an intra-assay imprecision of 2.0 - 3.0% coefficient of variation and an inter-assay imprecision of 3.2 - 4.4% coefficient of variation. The assay range is 0.23 - 7.25 mg/dL. A thorough review of different assay methodologies by Newman (2) concluded that it is the most precise automated assay across the clinical concentration range and performed slightly better than the other automated assays in terms of sensitivity and lack of analytical interference.
The sub-sample of participants aged 12-59 requires modification of original sampling weights. Weights will be provided by investigators to NCHS and will be adjusted by age, sex, race, and serum creatinine strata to match the original NHANES 1999-2000 design.
Acknowledgement
This research was supported by NIH grants UO1 DK 053869, UO1 DK 067651, and UO1 DK 35073.
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
0.47 to 7.15 | Range of Values | 2471 | 2471 | |
. | Missing | 37 | 2508 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
0 to 418199.9809 | Range of Values | 2508 | 2508 | |
. | Missing | 0 | 2508 |