June 20-21, 2006
Meeting Summary
|
|
Table of Contents
|
| I. Record of Attendance |
| II. Clinical Laboratory Improvement
Advisory Committee (CLIAC) - Background |
| III. Call to Order and Committee
Introductions
|
| IV. Agency Updates and Committee Discussion
|
| V. Presentations and Committee Discussion
|
| VI. Committee Discussion of Future Agenda
Items
|
| VII. Public Comments
|
VIII. Adjourn
|
| Record of Attendance |
|
| Committee Members Present
|
| Dr. Lou Turner, Chair |
|
Dr. Michael Laposata |
| Dr. Kimberle Chapin |
|
Dr. Dina Mody |
| Dr. Kathy Foucar |
|
Dr. Valerie Ng |
| Ms. Paula Garrott |
|
Dr. Barbara Robinson-Dunn |
| Dr. Lee Hilborne |
|
Dr. Jean Amos Wilson |
| Dr. Anthony Hui |
|
|
| Mr. Kevin Kandalaft |
|
|
|
|
| Committee Members Absent
|
| Ms. Joeline Davidson |
|
Dr. Patrick Keenan |
| Ms. Merilyn Frances |
|
Dr. Jared Schwartz |
| Dr. Peter Gomatos |
|
Dr. David Smalley |
| Dr. Carol Greene |
|
Dr. Thomas Williams |
| Ms. Luann Ochs, Roche Diagnostics Corporation (Liaison Representative - AdvaMed) |
|
|
| Executive Secretary
|
|
|
|
| Ex Officio Members |
| Dr. Thomas Hearn, CDC |
| Ms. Judith Yost, CMS |
| Ms. Carol Benson, FDA |
|
|
|
| Centers for Disease Control and Prevention
|
| Ms. Nancy Anderson |
|
Dr. Herschel Lawson |
| Ms. Pam Ayers |
|
Dr. James Little |
| Ms. Carol Bigelow |
|
Ms. Leslie McDonald |
| Ms. Diane Bosse |
|
Ms. Anne Pollock |
| Dr. Carlyn Collins |
|
Ms. Diane Ricotta |
| Ms. Carol Cook |
|
Ms. Andrea Scott |
| Ms. Stacey Cooke |
|
Mr. Howard Thompson |
| Ms. MariBeth Gagnon |
|
Ms. Pam Thompson |
| Mr. James Handsfield |
|
|
| Dr. Devery Howerton |
|
|
|
|
| Department of Health and Human Services
(Agencies other than CDC)
|
| Ms. Carol Benson (FDA) |
| Ms. Val Coppola (CMS) |
| Ms. Tremel Faison (FDA) |
| Ms. Louise Magruder (FDA) |
| Ms. Harriet Walsh (CMS) |
| Ms. Cheryl Wiseman (CMS) |
|
|
| In accordance with the provisions of Public Law 92-463, the meeting was open to
the public. Approximately 30 public citizens attended one or both days of the
meeting. |
| |
|
Clinical Laboratory Improvement Advisory Committee
|
|
The Secretary of Health and Human Services is authorized under Section 353 of
the Public Health Service Act, as amended, to establish standards to assure
consistent, accurate, and reliable test results by all clinical laboratories in
the United States. The Secretary is authorized under Section 222 to establish
advisory committees.
The Clinical Laboratory Improvement Advisory Committee (CLIAC) was chartered in
February 1992 to provide scientific and technical advice and guidance to the
Secretary and the Assistant Secretary for Health regarding the need for, and
the nature of, revisions to the standards under which clinical laboratories are
regulated; the impact on medical and laboratory practice of proposed revisions
to the standards; and the modification of the standards to accommodate
technological advances.
The Committee consists of 20 members, including the Chair. Members are selected
by the Secretary from authorities knowledgeable in the fields of microbiology,
immunology, chemistry, hematology, pathology, and representatives of medical
technology, public health, clinical practice, and consumers. In addition, CLIAC
includes three ex officio members, or designees: the Director, Centers for
Disease Control and Prevention; the Commissioner, Food and Drug Administration;
the Administrator, Centers for Medicare & Medicaid Services; and such
additional officers of the U.S. Government that the Secretary deems are
necessary for the Committee to effectively carry out its functions. CLIAC also
includes a non-voting liaison representative who is a member of AdvaMed and
such other non-voting liaison representatives that the Secretary deems are
necessary for the Committee to effectively carry out its functions. Due to the
diversity of its membership, CLIAC is at times divided in the guidance and
advice it offers to the Secretary. Even when all CLIAC members agree on a
specific recommendation, the Secretary may not follow their advice due to other
overriding concerns. Thus, while some of the actions recommended by CLIAC may
eventually result in changes to the regulations, the reader should not infer
that all of the Committee's recommendations will be automatically accepted and
acted upon by the Secretary.
|
|
CALL TO ORDER - INTRODUCTIONS/FINANCIAL DISCLOSURES
|
|
Dr. Lou Turner, Chair, Clinical Laboratory Improvement Advisory Committee (CLIAC), welcomed
the Committee members and called the meeting to order. All members then made self-introductions
and financial disclosure statements relevant to the meeting topics.
|
| |
|
AGENCY UPDATES AND COMMITTEE DISCUSSION
|
|
|
Centers for Disease Control and Prevention (CDC) Update
|
Dr. Robert Martin, Director, Division of Laboratory Systems (DLS [proposed]), National
Center for Preparedness, Detection, and Control of Infectious Disease (NCPDCID
[proposed]), Centers for Disease Control and Prevention (CDC), and Executive Secretary,
CLIAC, welcomed the members and requested their full attendance throughout the meeting
to ensure a quorum. He informed the Committee that DLS has completed the move to the
CDC Roybal Campus and Dr. Devery Howerton continues to serve as Acting Chief,
Laboratory Practice Standards Branch.
|
|
Centers for Medicare & Medicaid Services (CMS)
|
Addendum A
|
|
Ms. Judy Yost, Director, Division of Laboratory Services (DLS), Survey and
Certification Group (SCG), CMS, updated CLIAC on the status of proposed
legislation, Proficiency Testing Improvement Act of 2005 (H.R.4568), which has
passed the House and is now in the Senate Health, Education, Labor, and Pensions
(HELP) Committee. She stated CMS was asked to brief the HELP Committee on the
cytology proficiency testing (PT) regulations, the background for these regulations,
CMS' planned approach to regulatory changes, and the cytology PT data collected to
date. Ms. Yost indicated the HELP Committee intends to hold the bill so their
deliberations could include CLIAC's cytology PT recommendations.
Ms. Yost reviewed the regulatory process, including writing a preamble justifying the
proposed changes to the 1992 regulations, drafting the standards, conducting an
impact analysis, obtaining clearance from the agencies administering CLIA (CMS, CDC,
and, if needed, the Food and Drug Administration [FDA]), and providing response time
for public comment on the proposed rule. Ms. Yost expressed commitment from both CMS
and CDC to expedite the rule-making process once CLIAC's recommendations for cytology
PT regulatory revision are submitted to the Secretary, Department of Health and Human
Services (HHS). She stated the Committee would receive a progress report at the
September 2006 meeting.
Prior to presenting a brief overview of the preliminary 2005 national cytology PT results,
Ms. Yost explained to the Committee that testing initiated late in one calendar year
may carry over into the next year if retesting has not been completed, making data
analysis for a specific test year difficult. She further advised the members that,
while 11 laboratories had not enrolled in cytology PT by the end of 2005, all
individuals subject to cytology proficiency testing for 2005 were enrolled and tested
as of June 19, 2006. She summarized the 2005 data from the state of Maryland and
the Midwest Institute for Medical Education (MIME) programs including the overall
pass rate; a breakdown of pass rates for cytotechnologists, primary and secondary
screening pathologists, and locum tenens; and failure rates over time. In closing,
Ms. Yost encouraged use of the CMS web address (www.cms.hhs.gov/clia) for additional
information and provided the following list of CMS-approved cytology PT programs for
calendar year 2006:
- Maryland Cytology Proficiency Testing Program
- College of American Pathologists (CAP)
- American Society for Clinical Pathology (ASCP) (acquired the MIME program)
Committee Discussion
-
A member stated while CAP discourages primary screening by pathologists, many
pathologists might perform primary screening only rarely (as little as one day
a year). While not defending the practice of primary screening by pathologists,
this member described the required testing of these individuals as primary
screeners as a double burden.
-
An invited speaker commented that in April 2006 the American Society of
Cytopathology (ASC) adopted a position statement recommending that cytology
laboratories use cytotechnologists for primary screening of pap smears except
in emergency circumstances. Addendum
B
|
|
Regulatory Framework for Assuring Quality of Cytology Screening
|
Addendum
C*
|
|
Dr. Thomas Hearn, Associate Director for Laboratory Systems, DLS (proposed),
NCPDCID (proposed), CDC, thanked Dr. Diane Solomon for serving as Chair of the
Cytology Proficiency Testing Workgroup and the Workgroup members for their
efforts. He acknowledged the challenge confronted by the Workgroup and commended
the group for successfully putting aside individual interests to fulfill its
charge from CLIAC. After reviewing the relevant portions of the CLIA Law and
regulations, Dr. Hearn reminded the CLIAC members that implementation of the
regulations addressing quality assessment, personnel requirements, on-site
inspection, and other regulatory compliance activities have already improved
the quality of cytology testing.
In describing the chronology of events related to cytology PT and impediments
to the implementation of 1992 cytology PT regulations, Dr. Hearn discussed the
development of computer-based testing. This potential PT option was recommended by
CLIAC in 1993 in response to the reported lack of a sufficient quantity of glass
slides to implement a national cytology PT program. He summarized three CDC studies
to evaluate the efficacy of using virtual slides to replace glass slides, which
supported the use of computer-based testing in lieu of glass slide testing. Dr.
Hearn explained to the Committee that to replace glass slide testing with
computer-based testing would, however, require revisions to the 1992 CLIA regulations.
In addition, he identified other variables that can affect cytology PT outcomes. Dr.
Hearn concluded with a description of the collaborative nature of the rule-making
process and reminded the members that CLIAC's focus should remain on recommendations
to facilitate revisions to the 1992 CLIA cytology PT regulations.
*Note: The addendum was revised from material provided in the Committee's notebooks to reflect last minute updates by the presenter.
|
|
Cancer of the Cervix, An Overview
|
Addendum
D*
|
|
Herschel W. Lawson, M.D., Division of Cancer Prevention and Control, National Center for Chronic Disease Prevention and Health Promotion, CDC, presented information on the surveillance and incidence rates, screening, diagnosis, and associated costs of cervical cancer using data from the National Program of Cancer Registries and the Surveillance, Epidemiology, and End Results program. He pointed out that cervical cancer is the 11th most common cancer occurring in U.S. women, noting the most important factor contributing to cervical cancer is the absence or infrequency of receiving a screening test. In U.S. women, the incidence of cervical cancer has been reduced by 75% due to screening and early detection has increased survival rates. The estimated annual cost of screening and follow-up of abnormal Pap tests is more than $4 billion per year.
Committee Discussion
-
A member asked for clarification of the data comparing the 2002 and 2006 new cervical cancer cases versus numbers of deaths. He stated that while 2006 appeared to have significantly fewer new cases, there did not appear to be a corresponding decline in the death rate. In response, Dr. Lawson explained the 2002 data represent actual numbers while the 2006 figures represent American Cancer Society estimates, which will probably be higher when the 2006 data become available.
*Note: The addendum was revised from material provided in the Committee's notebooks to reflect last minute updates by the presenter.
|
|
Gynecologic Cytology Laboratory Practice and Methodology
|
Addendum
E
|
|
Dr. James V. Little, Assistant Professor of Pathology, Emory University School of
Medicine, and CDC consultant to DLS, gave an overview of pap smear testing and
cytology laboratory practices. He detailed the personnel skills required for
cytological examinations, cell classification terminology, and the role of pap smear
testing in the detection of gynecologic abnormalities and malignancies. Dr. Little
reminded the members PT is only one aspect of a cytopathology laboratory quality
system that includes quality control of staining and processing, personnel competency
evaluation, cytology-histology correlations, and random re-screening of normal smears,
as well as educational processes. He stated that despite low sensitivity the pap
test remains an effective means for cervical cancer detection. As such, it is most
effective when women are screened at regular, repeated intervals and when laboratory
practices are optimized. He ended his presentation by stating that the role of the
pap test in cervical cancer prevention is evolving.
Committee Discussion
-
Dr. Little was asked if data were available on the number
of abnormal smears detected because of the requirement to
re-screen 10 percent of normal smears. He explained data exist,
but the incidence rate of abnormal pap smears is sufficiently low
that a 10 percent re-screen is not the most effective tool for
identifying missed cases.
|
|
Cytology Proficiency Testing and New Technologies in Cervical Cancer Screening
|
Addendum
F
|
|
George G. Birdsong, M.D., Emory University School of Medicine and Grady Health System,
thanked TriPath Imaging and Cytyc Corporation for their assistance in developing his
presentation addressing cytology PT in the context of new technologies in cervical
cancer screening. He described the FDA-approved TriPath Focal Point Slide profiler
and Cytyc ThinPrep Imaging System, as well as the non-approved TriPath Focal Point
Guided Screening system; discussed how different technologies impact the cytology
screening process; and reviewed the three current sources of liquid-based pap tests
(Cytyc's ThinPrep, TriPath's SurePath, and MonoGen's MonoPrep). Dr. Birdsong also
briefly touched on the Digene Hybrid Capture 2, a molecular DNA test for 13 high-risk
types of human papillomavirus (HPV). He stated that FDA approval includes triaging
patients diagnosed with atypical squamous cells of undetermined significance (ASC-US)
and co-testing (use in conjunction with the pap smear) as routine screening of women
30 years of age or older. He concluded with a summary of issues associated with
attempting to apply cytology PT to new technologies. Automated screening presents
numerous challenges such as those arising from PT slide source variability, stain
variability, and morphologic differences due to the various preparation techniques.
The instrumentation, according to the manufacturers, is not ready to use for PT
because it does not allow repeat testing of the same slide or reading a slide on
multiple instruments.
Committee Discussion
-
A member inquired about screening the same slide on the same type of
instrument, but in different laboratories. Dr. Birdsong replied that
he was unaware of any peer-reviewed literature on this subject.
-
Dr. Turner asked if data were available on the percentage of conventional
versus liquid-based preparations for pap testing being performed. Dr.
Herschel Lawson stated the results of an independent CDC survey would be
published in the next few months in Obstetrics and Gynecology showing that
80-85 percent of all cytology testing is currently performed using liquid-based
specimens. A member commented that, while the conventional pap remains a
very good screening test, the option to perform ancillary testing on
liquid-based specimens has generated preference for this preparation among
gynecologists.
|
|
American Society of Clinical Pathology
|
Addendum
G*
|
|
Dr. Thomas Bonfiglio, PT Medical Director, ASCP, and Ms. Rhonda Metzler, PT Senior Manager,
ASCP, presented an overview of the structure of ASCP's CMS-approved cytology PT
program, summaries of test results, and proposals for regulatory changes. Dr.
Bonfiglio described the program as a collaborative effort to improve laboratory
quality and serve its members. Ms. Metzler detailed the ASCP program structure
including the proctor system, slide validation standards, and evaluation of testing
results. She also reviewed responses and rationales for test result appeals from
participants. Dr. Bonfiglio presented ASCP's recommendations for changes to the
regulations as follows: reduce the required testing interval from annual to every
two or three years supplemented with educational requirements in alternate years,
change the grading scheme to reflect current patient management practices, and
eliminate the automatic failure feature. He urged HHS to implement these changes
in 2006. Dr. Bonfiglio suggested that PT programs could also be improved by requiring
field validation of slides. He concluded with an update of the activities of the new
ASCP GYN PT & Assessment Committee.
*Note: The addendum was revised from material provided in the Committee's notebooks to reflect last minute updates by the presenter.
Committee Discussion
-
A member asked how it would be possible to misclassify a high grade
squamous intraepithelial lesion (HSIL) slide as negative in light of ASCP's
stringent validation requirements for test slides classified as HSIL, i.e.,
five reviewers must agree on the interpretation and histology/cytology
correlation must be documented. Dr. Bonfiglio stated cytology is subjective
and intra-observability variation makes it possible to misclassify a slide, but
ASCP hopes that by adding a requirement for 90 percent field validation by at
least 40 individuals misclassification will not be an issue. He noted that earlier
in the program, when slides were reviewed by three pathologists and field validation
was performed by the examinees, a few individuals misread HSIL slides. Dr.
Bonfiglio stressed the importance of increasing the number of slides per test
set and removing the possibility of an individual failing based on misinterpretation
of a single slide.
-
Another member registered concern over the proposed changes to the scoring
system that could result in failure to identify individuals interpreting pap
smears incorrectly. Without an integrated educational process, such individuals
could continue providing wrong diagnoses. The member suggested CLIAC recommend
individuals identified as failing a testing event should participate in targeted
educational processes and strive to achieve the highest attainable score.
Another member concurred, but added that patient and clinician education is also
needed, noting that 90 percent of cervical cancers occur in women who have either
never been screened or have not been screened in the past five years, or whose
pap smear is reported as abnormal but whose case is subsequently mismanaged; in
contrast, interpretive errors account for 5-10 percent of missed cases.
-
The same member expressed concern that the low number of PT appeals in 2005 could
be attributed to the fact that the first individuals tested were unaware they could
appeal their scores. Dr. Bonfiglio explained the MIME program (now ASCP) built
checks into its PT program to address this issue. When an individual failed PT,
the slide was pulled and subjected to a blinded review immediately. This de facto
appeal process resulted in the PT program removing slides that had fallen out of
validation and reclassifying individuals who failed based on such slides to passing
status. Every slide in the program has now been validated at the 90% level with a
minimum of 40 reviews.
-
Dr. Turner requested and received clarification that individuals employed in
more than one laboratory are required to undergo PT at only one site.
-
One member noted the PT result data show more individuals choose to discontinue
testing rather than attempt a second or third test and asked what is known about
that group. Dr. Bonfiglio responded that the history of cytology PT demonstrates
few examinees truly fail, while a greater number of individuals stop performing
gynecologic cytology after failing the first test set and are no longer required to
participate in the examination process. He further stated that when evaluating
cytology PT effectiveness, it is important to consider not only the number of
failures but also the number of individuals who have decided to cease PT and screening.
|
|
The Maryland Cytology Proficiency Testing Program
|
Addendum H
|
|
The state of Maryland submitted a written overview of the Maryland Cytology
Proficiency Testing Program for CLIAC's review.
|
|
The New York State Department of Health Cytopathology Proficiency Testing Program: Lessons Learned and Recommendations
|
Addendum
I
|
|
Ms. Deirdre Astin, Deputy Director of the Clinical Laboratory Evaluation Program
at the New York State Department of Health, Wadsworth Center, began her presentation
by providing the background and history of the New York State Department of Health
Cytopathology Proficiency Testing Program. She stated the original state PT program
was mandated in 1964, and an annual on-site glass slide evaluation program was
established in 1968. Ms. Astin talked about the issues the program faced after
changing to a biennial testing format in 1978 and then compared test design and
grading systems of the New York State cytology proficiency program with those of
CLIA. She compared performance statistics of the New York State program with the
national statistics of the MIME program and concluded by discussing observations on
test performance and considerations for CLIAC to include in their discussion on
cytology proficiency testing.
|
|
Limitations of the Ten Slide Proficiency Test Model
|
Addendum
J
|
|
Dr. George K. Nagy, Section Head, Pathology, New York State Department of Health,
Wadsworth Center, began by stating the current ten-slide cytology proficiency testing
model lacks the discriminatory power to differentiate between competent and incompetent
test takers. He showed how the ten-slide test could misclassify test takers, with a
high proportion of competent individuals failing and a high proportion of incompetent
individuals passing. He went on to say that a competent examinee failing a test
event (type 1 error) is the lesser problem because of the high probability that the
individual would pass the second test event. However, an incompetent examinee
passing the test event (type 2 error) is a more serious problem as this individual
will continue to screen patient specimens until the next test cycle. Using statistical
examples, Dr. Nagy demonstrated how larger sample sizes provide greater discrimination
in probability distribution with more accurate estimations. Therefore, use of large
samples in PT could substantially reduce misclassifications, especially those arising
from type 2 errors. He proposed the main objectives of cytology PT should be the
detection and re-education of incompetent individuals followed by retesting to assure
competency. In summary, Dr. Nagy proposed a more accurate assessment of competency
would be an initial test consisting of 40-60 slides administered upon entry of
pathologists and cytotechnologists into practice. Additionally, he suggested the
initial assessment be followed by PT at five- to ten-year intervals as is currently
the practice for recertification by many professional boards since the ability to
screen slides does not diminish over time.
|
|
Cytology Proficiency Testing Workgroup Report
|
Addendum
K*
|
|
Dr. Diane Solomon, Senior Investigator, Division of Cancer Prevention,
National Cancer Institute, National Institutes of Health, DHHS, provided
CLIAC with a summary of proceedings from the Cytology Proficiency Testing
Workgroup meeting. She began by restating the Workgroup's charge: "Openly
discuss issues, consider comments and develop potential framework and options
for regulatory revisions that will be reported to CLIAC for developing
recommendations to HHS for assisting in the development of a proposed rule."
Dr. Solomon informed the members the list of issues presented for the
Workgroup's consideration was derived from organizational comments and
presented the Workgroup's comments and suggested options for regulatory
revision for each of the following cytology PT issues:
- individual vs. laboratory
- new technology
- testing frequency
- number of challenges
- categories of challenges
- number of challenges per category
- grading scheme
- validation
- test site
- retesting
- confidentiality
Dr. Solomon closed by summarizing the various options for regulatory revisions
developed by the Workgroup for CLIAC's consideration.
*Note: The addendum was revised from material provided in the
Committee's notebooks to reflect last minute updates by the presenter.
|
|
Analysis of Data from the Maryland Cytology Proficiency Testing Program
|
Addendum
L*
|
|
Dr. Devery Howerton, Chief, Laboratory Practice Evaluation and Genomics Branch, and Acting Chief,
Laboratory Practice Standards Branch, DLS (proposed), NCPDCID (proposed), CDC,
prefaced her presentation by clarifying for CLIAC the cytology PT changes that
could be achieved through regulatory revision versus those requiring statutory
change. She explained her objective was to use raw data from the Maryland Cytology
Proficiency Testing Program to present examples using the scoring models proposed
by the Workgroup. This could provide a scientific basis for scoring changes that
CLIAC might recommend. She then summarized the outcomes from analyses of 1998-2004
raw data from the Maryland program, first by evaluating performance according to the
CLIA grading scheme and then by comparing four alternate grading schemes to CLIA.
Next, she compared the results of a simulated 20-slide test with the 10-slide test,
emphasizing this did not represent a true comparison study and acknowledging the
results were thus inconclusive. Dr. Howerton also noted the Maryland data were
inadequate for evaluating test frequency. With respect to frequency of testing,
however, she commented that a retrospective analysis of the New York cytology PT
program performance data showed no significant change in failure rate following a
change from annual to biennial testing. Dr. Howerton concluded by again identifying
the proposed cytology PT options provided by the Workgroup for discussion and
acknowledged the individuals providing analysis of the Maryland raw data.
*Note: The addendum was revised from material provided in the Committee's notebooks to reflect
last minute updates by the presenter.
Committee Discussion: Cytology Proficiency Testing
The CLIAC Chair introduced Dr. Solomon as technical moderator for the CLIAC
discussions and referred the Committee to the list of the cytology PT issues to
be considered. Several Committee members sought clarification on what specific
elements of cytology PT could be changed by revision of the CLIA cytology PT
regulations. Dr. Howerton reiterated changes in number and category of challenges
and frequency of testing could be addressed by regulatory revision but PT of
individuals is required by the CLIA statute and could not be addressed via this
process.
Dr. Solomon reviewed the Workgroup's discussion and asked CLIAC to deliberate and
make recommendations based on the Workgroup's options for regulatory revisions. The
following represents a per-issue summary of options for regulatory revision
identified by the Workgroup, the Committee's deliberations, and final CLIAC
recommendations.
Note: Questions intended to stimulate CLIAC's consideration of each issue were provided
as part of the Committee reference materials, but were not specifically used to guide
the members' discussion. (Addendum M)
Individual vs. Laboratory
CLIAC members expressed concern that the statute required PT of individuals only in
the subspecialty of cytology and stated this was unfair and should be an issue open
for discussion. Several members also suggested CLIAC consider PT proposals
presented in the PT providers' public comments. Dr. Turner reminded the Committee
the objective of this meeting was to focus on the Workgroup report and develop
recommendations addressing cytology PT requirements subject to change by regulatory
revision. She reemphasized the issue of individual versus laboratory PT could not
be addressed via this process. Before the Chair closed discussion of this topic,
several members proposed the preamble language should reflect strong encouragement
for laboratory participation in cytopathology educational programs in addition to
individual proficiency testing. The Committee agreed, adding the CLIA interpretive
guidelines could also be used to encourage oversight organizations/agencies and
surveyors to determine if laboratories are participating in educational programs
during on-site inspections. If non-participation is discovered then the site
surveyor could identify available resources and encourage laboratories to enroll
in cytopathology educational programs.
CLIAC Recommendations
- Use the preamble to encourage laboratories to participate in educational
laboratory programs in addition to individual proficiency testing
- Oversight organizations/agencies and surveyors should
determine if laboratories participate in educational programs
and help to identify available resources if they are not
New Technology
One member suggested the regulatory language of "glass slide" be replaced with
terminology that more realistically reflects the present and the future of
gynecological testing. Another member cautioned the replacement language should
not promote a specific technology nor inhibit the introduction of new technologies.
The Committee agreed and recommended the regulatory language of "challenges"
replace "slides" and challenges be defined as "case equivalent - glass slide,
virtual slide, or other approved media." Several members also commented the use
of "challenges" would allow flexibility to both PT providers in transitioning
from using older technologies to newer technologies and to individuals in
selecting a technology offering a greater comfort level. The CLIAC also
recommended adding a regulatory requirement establishing a transition phase when
new formats are introduced to allow individuals their choice of format for all
retesting phases.
CLIAC Recommendations
- Change current language of "slides" to "challenges" to allow for the use of virtual slides or other media
- Define a challenge as a case equivalent - glass slide, virtual slide, or other approved media
- Add requirement for a transition phase for new technology (e.g., virtual slides), when the individual can request retesting with previous platform/format (e.g., glass slides)
Frequency of Testing and Number of Challenges
Dr. Solomon introduced the issues of test frequency and number of
challenges, noting the two issues cannot be separated, since the Workgroup
linked specific options for revisions to whether the number of challenges was
increased concomitant with a decrease in frequency. The Workgroup favored a
20-challenge test every two-plus years. Dr. Solomon clarified the decreased
frequency is justified since there is general agreement in the profession that
individual skills do not deteriorate over time. In addition, she explained an
increase in the number of challenges would decrease the effectiveness of gaming
tactics. A member suggested the increased demands associated with preparing
20-challenge test sets should be offset in part by less frequent testing.
Regarding the number of challenges and testing frequency, Dr. Solomon commented
on Dr Nagy's proposal of 40- to 60-slide tests, saying the suggestion is
theoretically appealing but testing twice with 20 slides within a six-year
period would have the same mathematical effect as the larger test every six
to ten years. Several Committee members commented that 40- to 60-slide tests
could pose significant logistical challenges both for vendors and in terms of
personnel coverage during time away from the bench. The Committee agreed to
recommend 20 challenges for every test and to reduce the frequency of testing
to a 3-year test cycle. PT providers noted initial testing of new personnel
could still be accommodated as needed while routine testing in laboratories
could be spread out over the three years.
CLIAC Recommendations
- Reduce the frequency of testing to a 3-year test cycle
- Use 20 challenges for every test (initial test and retests)
Categories of Challenges and Number of Challenges per Category
The Committee agreed with the Workgroup's proposed option of at least one
challenge from each category based on a 20-challenge test.
CLIAC Recommendation
- Retain four response categories and continue to require at least one
challenge from each of the four categories in each test
Grading Scheme
Prior to discussing specific workgroup comments and suggestions regarding
the grading scheme, Dr. Solomon advised the Committee of the workgroup's
approach to cytology PT grading issues. She explained their intent to propose
a simplified, more valid grading system with response categories and scoring
that reflect clinical practice decision points in patient management. Dr.
Solomon further explained the workgroup's efforts to remain neutral in terms
of affecting the current pass/fail rates throughout the process.
Dr. Solomon advised the Committee that the Workgroup strongly supported
removing the automatic failure requirement (HSIL misinterpreted as negative
resulting in a 15-point loss), explaining failure based on a single slide was
deemed too punitive. In addition, the Workgroup supported eliminating partial
credit for identifying a low grade squamous intraepithelial lesion (LSIL) as
unsatisfactory (UNSAT). Dr. Solomon clarified the Workgroup's position that
individuals should be able to distinguish between LSIL and UNSAT. She further
noted elimination of partial credit would minimize examinees' ability to game
the test system. However, misinterpretation of a "negative" as "unsatisfactory"
should receive partial credit because in practice unsatisfactory slides result
in the patient getting a repeat pap smear.
Dr. Solomon briefly reviewed cytological interpretation and clinical
management models for the respective diagnoses of LSIL and HSIL, which differ
from the standard of practice in 1992 when the current CLIA cytology PT regulations
were finalized. She said the Workgroup was divided on whether to retain the
requirement to distinguish between LSIL and HSIL.
Dr. Solomon told the Committee the Workgroup was also divided on whether there
should be a unified or separate scoring system for pathologists and cytotechnologists
and presented arguments supporting both approaches for consideration.
In favor of separate scoring systems:
- The cytotechnologist's primary responsibility is to alert the pathologist of
a possible abnormality and refer cases, not to diagnose specific abnormalities.
- Responsibility for the final interpretation rests with the pathologist, not the cytotechnologist.
In support of a unified scoring system:
-
Cytopathologists and cytotechnologists receive the same training and
education in morphologic criteria, e.g., diagnosis of HSIL versus LSIL
and normal versus unsatisfactory.
-
Although the pathologist is ultimately responsible for interpretation
of abnormals, in most cases the cytotechnologist's interpretation becomes
the final diagnosis.
Committee Discussion
- Regarding the Workgroup's suggested removal of the automatic failure requirement, several members commented on the need to be sensitive to public perception when determining the scoring weight given to misinterpretation of HSIL as negative.
- In response to a member's question concerning how many examinees failed PT based solely on an automatic failure (misinterpreted an HSIL as negative and correctly interpreted the other nine test slides), Dr. Howerton responded automatic failure occurred with 11% of cytotechnologists and 7% of pathologists based on the Maryland program's data.
- A member noted that, for a 20-slide test, if the -5 was retained, an examinee misinterpreting one HSIL as negative would lose a total of 10 points and would have to provide completely correct responses to the remaining 19 slides in order to avoid a retest. Another member pointed out a 20-slide test might easily contain more than one HSIL slide, which could help discern whether an individual truly needs remediation in this area. A member moved to retain the -5 scoring for misinterpreting HSIL as negative, resulting in a significant penalty commensurate with the seriousness of the error but not in an automatic failure when using a 20-challenge test.
- Some committee members agreed that in many practices, pathologists "rubber stamp" the cytotechnologist's interpretation. In addition, if the cytotechnologist does not dot the cells, pathologists may not see the cells or re-screen the slides. A PT provider noted, in the previous year's testing, that pathologists provided the same response as their cytotechnologist screeners 90% of the time.
- A member emphasized the ultimate responsibility should fall on the decision maker, which in this case is the pathologist.
- Another member stated cytology PT holds examinees to a higher standard than clinical practice in that the latter allows the pathologist to include qualifying comments in the final interpretation.
- Dr. Hearn pointed out that 90% of the slides that are screened are interpreted as normal and
never seen by a cytopathologist. Therefore, cytotechnologists have a high level of responsibility for reporting normal results and cytopathologists for reporting abnormal results. In principle, one could argue there is a balance in these responsibilities that might support a unified scoring scheme.
- Several Committee members commented that changes to PT should not additionally burden the cytotechnologists in comparison to current CLIA scoring, however the proposed unified system is slightly less rigorous for pathologists and slightly more challenging to cytotechnologists.
- A member proposed a unified scoring system with no penalty for distinguishing between LSIL and HSIL for pathologists.
Following extensive deliberations as summarized above, CLIAC agreed to recommend a unified scoring system that eliminates the automatic failure component yet imposes a significant penalty for misinterpretation of HSIL as negative. In addition, the recommended scoring system minimizes the distinction between LSIL and HSIL by requiring examinees to distinguish between these diagnoses but with no loss of points.
CLIAC Recommendation
-
Change grading scheme to Unified Model X as shown below
|
|
Model X-20-Slide Test - Unified
|
| Correct Response |
Examinee Response |
| |
A - UNSAT |
B-NEGATIVE |
C - LSIL |
D - HSIL |
| A - UNSAT |
5 |
0 |
0 |
0 |
| B - NEGATIVE |
2.5 |
5 |
0 |
0 |
| C - LSIL |
0 |
0 |
5 |
5 |
| D - HSIL |
0 |
-5 |
5 |
5 |
Validation and Test Result Feedback
Dr. Solomon explained the reasoning behind the Workgroup's
proposal to drop the requirement for biopsy confirmation of
slides referenced as LSIL. While LSIL is a reproducible diagnosis,
there are instances of cytologic LSIL that do not confirm colposcopically
because colposcopy is not highly sensitive and these lesions are often
transient and frequently regress in the interval between the time the pap
smear is taken and the time of colposcopic biopsy. Therefore, the Workgroup
supported dropping the requirement for biopsy confirmation of LSIL, thus
making it easier for programs to collect examples of these cases, assuming
appropriate slide validation.
The Committee discussed making field validation and disclosure of PT providers' validation
process a regulatory requirement as follows:
- The issue of a slide gradually fading was raised as a factor in field validation. Additionally, PT providers noted liquid-based preparations tend to fade more rapidly than conventional smears.
- Specifically addressing the concern an HSIL slide may fall out of validation, one PT provider stated that every failure goes through a de facto appeal process, which includes technical evaluation, before results are reported.
- Another PT provider agreed that every failure for individuals is reviewed prior to results being returned to participants. Additionally, review of every slide when it is returned to the provider offers the opportunity to remove slides when the quality deteriorates. If an appeal is upheld or if a slide is removed after review, the grades are changed.
- Dr. Turner confirmed that when slides do not validate upon return the current PT providers adjust PT scores accordingly and notify affected examinees if their adjusted score results in a change to their pass/fail status.
- One PT provider responded that each time test sets are returned, the provider evaluates every slide as well as the participants' performance and applies algorithms to determine fair and equitable test set composition for the next round.
The Committee discussed recommending PT providers give examinees educational feedback of their test performance beyond the current requirement of informing examinees only of their area of test failure, noting such practice was already incorporated into cytology education-only programs.
- The Committee asked PT providers to comment on their ability to provide individual test scoring feedback without compromising test security. One provider cited no problems with test security because their slide set composition is determined for each testing event by application of computer algorithms to a large reservoir of slides. Another provider cited a potential security problem since slides in every test set were not always changed out before redistribution.
- A member suggested feedback be given only for discrepant answers.
- While stating it would be ideal to derive as much educational value as possible from each proficiency test, Dr. Solomon also commented on the additional burden to vendors if they are required to design cytology proficiency tests that function additionally as educational products, especially since participation in educational programs is already being encouraged.
- In response to a member's question, Ms. Yost reminded the Committee that CLIA prohibits examinee discussion or sharing of testing information prior to completion of the testing event.
- Several members suggested vendors furnish detailed feedback including examinee response, correct diagnosis, and any discrepancies. To prevent fraudulent sharing of results, the slides would not need to be specifically identified.
- Ms. Yost stated ideally cytology PT should correlate with clinical laboratory PT with respect to provider feedback and the required laboratory response. Regardless of whether a passing score is achieved, the laboratory is expected to review every incorrect response and to investigate the reason for any errors as a part of ongoing quality assessment.
CLIAC Recommendations
- Require biopsy confirmation of category D (HSIL/cancer) challenges, but not category C (LSIL) challenges
- Require field validation, monitor challenges continuously, and remove challenges that fail field validation
- Require validation procedures to be disclosed by the vendor
- Provide educational feedback for discrepancies
Test Site
Several members stated that laboratory directors should have the option to designate proctors. The Committee was informed that PT providers already determine the proctor requirements as well as retesting sites and was advised that specifying proctor requirements in the regulation could potentially hinder introduction of new technology or other proficiency testing formats.
CLIAC Recommendation
PT providers should continue to determine proctor requirements.
Retesting
Regarding the issue of retesting, the members agreed with the Workgroup's options for regulatory revision.
(Addendum K)
CLIAC Recommendations
- Require PT providers to disclose the appeal process in writing
- Change language to state "individuals who score <90% must."(as opposed to "who fail")
Confidentiality
The Workgroup requested clarification from CMS regarding the issues of confidentiality of test scores.
Dr. Solomon asked Ms. Yost to describe how CMS addresses confidentiality issues. Ms. Yost informed CLIAC of an
informational supplement on the CMS website
(www.cms.hhs.gov/CLIA/downloads/Informational_Supplement.pdf) addressing frequently
asked questions on cytology PT, including confidentiality issues and the release of
individual test performance data. She encouraged CLIAC members to visit the CMS
website to review the information provided. She reminded CLIAC the years 2005 and
2006 were educational cytology PT cycles therefore no individual scores are being
reported to CMS. She described the CMS Cytology Personnel Records System (CYPERS),
created in 2005, as a confidential records system linking laboratory demographics and
cytology testing personnel. She indicated CMS would be addressing record retention
issues in collaboration with CDC and agency lawyers this year. Additionally, she
assured CLIAC that CMS protocol requires case-by-case evaluation for response to
Freedom of Information Act (FOIA) requests for individual information and stated,
in her experience, most litigation and subpoenas issued for release of information
are directed to the individual, laboratory, and/or laboratory director and not to
CMS.
Committee Discussion
- One member opened the discussion by asking for clarification of what individual PT performance information should be kept confidential. This member expressed concern that individuals failing all phases of remediation could go unrecognized. Ms. Yost responded all information in CYPERS is confidential and PT providers would not release scores without written consent from that individual.
- Several members expressed concern that individuals failing four rounds could be hired by another laboratory. Guidance to laboratory directors was recommended, advising the necessity to conduct primary source verification (i.e., obtain prior written authorization from the examinee to the PT provider for release of test results directly to the prospective employer) of prospective employees' PT status prior to hiring. Ms. Yost pointed out much of this concern should be alleviated by the fact that oversight organizations/agencies and CMS surveyors will have information, at least at the laboratory level, on individuals who have not passed.
CLIAC Recommendation
None
Note: CLIAC recommendations captured during the meeting are reflected in Addendum N.
Other Committee discussion
The Committee suggested reviewing and evaluating PT program performance for the next several years to decide if further scoring modifications are necessary.
A member asked when the PT regulation would be published and implemented. Ms. Yost told the Committee it is on the CMS regulatory agenda and reviewed the steps for moving a proposed regulation through the clearance and publication process.
|
PUBLIC COMMENTS
|
|
Maryland Cytology Proficiency Testing Program
|
Addendum O
|
|
Ms. Barbara Keller, Maryland Cytology Proficiency Testing Program
|
Addendum P
|
|
Ms. Janie Roberson, American Society for Cytotechnology
|
Addendum Q
|
|
The College of American Pathologists
|
Addendum R
|
|
Dr. Thomas Wheeler, The College of American Pathologists
|
Addendum S
|
ADJOURN
Dr. Turner again acknowledged Dr. Solomon and the Cytology PT Workgroup and thanked the CLIAC members and partner agencies for their support and participation. The following reflects CLIAC's recommendations for revisions to the CLIA cytology PT regulations:
- Use the preamble to encourage laboratories to participate in educational laboratory programs in addition to individual proficiency testing
- Oversight organizations/agencies and surveyors should determine if laboratories participate in educational programs and help to identify available resources if they do not
- Change current language of "slides" to "challenges" to allow for the use of virtual slides or other media
- Define a challenge as a case equivalent - glass slide, virtual slide, or other approved media
- Add requirement for a transition phase for new technology (e.g., virtual slides), when the individual can request retesting with previous platform/format (e.g., glass slides)
- Reduce the frequency of testing to a 3-year test cycle
- Use 20 challenges for every test (initial and retest)
- Retain four response categories and continue to require at least one challenge from each of the four categories in each test set
- Change grading scheme to a new model that is the same for both technical supervisors and cytotechnologists
|
|
Model X-20-Slide Test - Unified
|
| Correct Response |
Examinee Response |
| |
A - UNSAT |
B-NEGATIVE |
C - LSIL |
D - HSIL |
| A - UNSAT |
5 |
0 |
0 |
0 |
| B - NEGATIVE |
2.5 |
5 |
0 |
0 |
| C - LSIL |
0 |
0 |
5 |
5 |
| D - HSIL |
0 |
-5 |
5 |
5 |
- Require biopsy confirmation of category D (HSIL/cancer) challenges, but not category C (LSIL) challenges
- Require field validation, monitor challenges continuously, and remove challenges that fail field validation
- Require validation procedures to be disclosed by the vendor
- Provide educational feedback on discrepancies
- Require PT providers to disclose the appeal process in writing
- Change language to state "individuals who score <90% must."(as opposed to "who fail")
Dr. Turner announced the next CLIAC meeting is scheduled for September 20-21, 2006, and adjourned the Committee meeting.
I certify this summary report of the June 20-21, 2006, meeting of the Clinical Laboratory Improvement Advisory Committee is an accurate and correct representation of the meeting.
| Lou Flippin Turner, Dr.P.H., CLIAC Chair |
Dated: 9/7/2006 |
|
This page last reviewed: 9/22/2006
|
|