Component Description

Perchlorate, Nitrate, and Iodide in Water (Home Tap Water):

By assessing exposure to these three toxicologically-related analytes (perchlorate, nitrate and iodide) in one assay, the relative impact of each chemical on thyroid function can be estimated and thus provide useful information for assessing the potential association between exposure and health effects.

Eligible Sample

Participants aged 12 years and older on a one-half sample had water specimens tested.

Description of Laboratory Methodology

Measurements of Perchlorate, Nitrate, and Iodide in Tap Water:

This method is a quantitative procedure for the measurement of perchlorate, nitrate and iodide in tap water using ion chromatography coupled with electrospray tandem mass spectrometry.

Chromatographic separation is achieved using an IonPac AS20 column with sodium hydroxide as the eluant. The eluant from the column is ionized using an electrospray interface to generate and transmit negative ions into the mass spectrometer. Comparison of relative response factors (ratio of native analyte to stable isotope-labeled internal standard) of unknowns with known standard concentrations yields individual analyte concentrations.

Data Processing and Editing

Automated data collection procedures for the survey were introduced in NHANES 2005–2006. In the MECs and analytical laboratories, data for the laboratory component are recorded directly onto a computerized data collection form. The system is centrally integrated, and it allows for ongoing monitoring of much of the data. Although the complete blood count and pregnancy analyses are performed in the MEC laboratory, most analyses are conducted elsewhere by approximately 28 laboratories across the United States.

Guidelines have been developed that provide standards for naming variables, filling missing values, and handling missing records. NCHS staff, assisted by contract staff, has developed data-editing specifications that check data sets for valid codes, ranges, and skip-pattern consistencies and examine the consistency of values between interrelated variables. Comments have been reviewed and recoded. NCHS staff verifies extremely high and low values whenever possible, and numerous consistency checks are performed. Nonetheless, users should examine the range and frequency of values before analyzing data.

Data Editing

The data editing specifications are as follows:

  • Age and gender checks
  • Total number of observations complete for each field
  • No field overlap, truncated values, or atypical results
  • Direct data entry errors
  • Abnormal results confirmed by lab
  • Test algorithm performed
  • Checked comment codes to resolve missing results and missing records
  • All missing results and missing MEC-examined records are accounted
  • Duplicate records were verified and deleted

Laboratory Quality Assurance and Monitoring

Mobile Examination Centers (MECs)

Laboratory team performance is monitored using several techniques. NCHS and contract consultants use a structured quality assurance evaluation during unscheduled visits to evaluate both the quality of the laboratory work and the quality-control procedures. Each laboratory staff person is observed for equipment operation, specimen collection and preparation; testing procedures and constructive feedback are given to each staff. Formal retraining sessions are conducted annually to ensure that required skill levels were maintained.

The NHANES QA/QC protocols meet the 1988 Clinical Laboratory Improvement Act mandates. Detailed QA/QC instructions are discussed in the NHANES LPM.

Analytical Laboratories

NHANES uses several methods to monitor the quality of the analyses performed by the contract laboratories. In the MEC, these methods include performing blind split samples collected on “dry run” sessions.

NCHS developed and distributed a quality control protocol for all the contract laboratories which outlined the Westgaard rules used when running NHANES specimens. Progress reports containing any problems encountered during shipping or receipt of specimens, summary statistics for each control pool, QC graphs, instrument calibration, reagents, and any special considerations are submitted to NCHS and Westat quarterly. The reports are reviewed for trends or shifts in the data. The laboratories are required to explain any identified areas of concern.

NCHS/Westat is currently reviewing these reports.

Analytic Notes

Measures of perchlorate, nitrate, and iodide in water were assessed in a subsample of participants aged 12 years and older. Use the special weights (WTSPC2YR) included in this data file when analyzing data. Please refer to the Analytic Guidelines for further details on the use of sample weights and other analytic issues.

The analysis of NHANES 2005-2006 laboratory data must be conducted with the key survey design and basic demographic variables. The recommended procedure for variance estimation requires use of stratum and PSU variables (SDMVSTRA and SDMVPSU, respectively), which are included in the demographic data file for each data release. The NHANES 2005-2006 Household Questionnaire and Demographic Data Files contain demographic data, health indicators, and other related information collected during household interviews. The phlebotomy file includes auxiliary information such as the conditions precluding venipuncture. The demographic, household questionnaire and phlebotomy files may be linked to the laboratory data file using the unique survey participant identifier SEQN.

Detection limits

The detection limit was constant for the analytes in the data set. In addition two variables are provided for each of these analytes. The variable named LBD___LC indicates whether the result was below the limit of detection. There are three values: “0”, “1”, and “2”. “0” means that the result was at or above the limit of detection. “1” indicates that the result was below the limit of detection. “2” means the result was above the limit of detection and exceeds calibrated range of assay.

The other variable named LBX___ provides the analytic result for that analyte. In cases, where the result was below the limit of detection, the value for that variable is the detection limit divided by the square root of two.

References

Codebook and Frequencies

SEQN - Respondent sequence number

Variable Name:
SEQN
SAS Label:
Respondent sequence number
English Text:
Respondent sequence number.
Target:
Both males and females 12 YEARS - 150 YEARS

WTSPC2YR - Two-year MEC weight of water perchlorate

Variable Name:
WTSPC2YR
SAS Label:
Two-year MEC weight of water perchlorate
English Text:
Two-year MEC weights of water perchlorate
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0 to 386575.00873 Range of Values 3435 3435
. Missing 0 3435

LBXWIO - Iodide, water (ng/mL)

Variable Name:
LBXWIO
SAS Label:
Iodide, water (ng/mL)
English Text:
Iodide, water (ng/mL)
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0.14 to 904 Range of Values 3262 3262
. Missing 173 3435

LBDWIOLC - Iodide, water comment code

Variable Name:
LBDWIOLC
SAS Label:
Iodide, water comment code
English Text:
Iodide, water comment code
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0 At or above the detection limit 3212 3212
1 Below lower detection limit 50 3262
. Missing 173 3435

LBXWNO - Nitrate, water (ng/mL)

Variable Name:
LBXWNO
SAS Label:
Nitrate, water (ng/mL)
English Text:
Nitrate, water (ng/mL)
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
494.97 to 80500 Range of Values 3261 3261
. Missing 174 3435

LBDWNOLC - Nitrate, water comment code

Variable Name:
LBDWNOLC
SAS Label:
Nitrate, water comment code
English Text:
Nitrate, water comment code
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0 At or above the detection limit 1782 1782
1 Below lower detection limit 1479 3261
. Missing 174 3435

LBXWP8 - Perchlorate, water (ng/mL)

Variable Name:
LBXWP8
SAS Label:
Perchlorate, water (ng/mL)
English Text:
Perchlorate, water (ng/mL)
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0.07 to 18.2 Range of Values 3262 3262
. Missing 173 3435

LBDWP8LC - Perchlorate, water comment code

Variable Name:
LBDWP8LC
SAS Label:
Perchlorate, water comment code
English Text:
Perchlorate, water comment code
Target:
Both males and females 12 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0 At or above the detection limit 2793 2793
1 Below lower detection limit 469 3262
. Missing 173 3435