Mercury is widespread in the environment and originates from natural and anthropogenic sources. The general population may be exposed to three forms of mercury: elemental, inorganic, or organic (primarily methylmercury). The concentration of total mercury, in urine, is a biomeasure of exposure primarily to elemental and inorganic mercury. Elemental and inorganic mercury exposure can result from mercury spills, dental amalgams, and occupational exposures. Both elemental and inorganic mercury are nephrotoxic and neurotoxic. Health effects related to low exposure in the general population are not well defined. In the 1999-2002 NHANES, urine mercury levels were measured in all women aged 16-49 years. From 2003-2004 forward, urine mercury levels were measured in a one-third subsample of participants aged 6 years and older.

Eligible Sample

Participants aged 6 years and older, on a one-third sample, were eligible.

Description of Laboratory Methodology

Urine iodine and mercury concentrations are determined by ICP-DRC-MS (Inductively Coupled Plasma Dynamic Reaction Cell Mass Spectroscopy). This multielement analytical technique is based on quadrupole ICP-MS technology and includes DRC™ technology. Coupling radio frequency power into a flowing argon stream(seeded with electrons) creates the plasma, the heat source, which is ionized gas suspended in a magnetic field. Predominant species in the plasma are positive argon ions and electrons. Diluted urine samples are converted into an aerosol using a nebulizer inserted within the spray chamber. A portion of the aerosol is transported through the spray chamber and then through the central channel of the plasma, where it is exposed to temperatures of 6000-8000 ºK. This thermal energy atomizes and ionizes the sample. The ions and the argon enter the mass spectrometer through an interface that separates the ICP, which is operating at atmospheric pressure (approximately 760 torr), from the mass spectrometer, which is operating at approximately 10-5 torr. The mass spectrometer permits detection of ions at each mass-to-charge ratio in rapid sequence, which allows the determination of individual isotopes of an element. Once inside the mass spectrometer, the ions pass through the ion optics, then through DRC™, and finally through the mass-analyzing quadrupole before being detected as they strike the surface of the detector. The ion optics uses an electrical field to focus the ion beam into the DRC™. The DRC™ component is pressurized with an appropriate reaction gas and contains a quadrupole. Electrical signals resulting from the detection of the ions are processed into digital information that is used to indicate the intensity of the ions and subsequently the concentration of the element. Traditionally ICP-MS has been a trace analysis technique and the typical measurement ranges from 0.1µg/L to around 100 µg/L. DRC technology provides additional control of ICP-MS sensitivity; therefore appropriate adjustments of the reaction cell parameters can significantly extend the useful concentration measurement range. In this method, iodine (isotope mass 127), tellurium (isotope mass 130), mercury (isotope mass 202) and bismuth (isotope mass 209) are measured in urine by ICP-DRC-MS using 100% argon as the Dynamic Reaction Cell™ (DRC) gas utilizing collisional focusing. Urine samples are diluted 1+1+ 8 (sample+ water + diluent) with water and diluent containing tellurium and bismuth for internal standardization.

Laboratory Quality Assurance and Monitoring

Specimens were processed, stored and shipped to Division of Laboratory Sciences, National Center for Environmental Health, National Centers for Disease Control and Prevention, Atlanta, Georgia.

The NHANES quality control and quality assurance protocols (QA/QC) meet the 1988 Clinical Laboratory Improvement Amendments mandates. Detailed quality control and quality assurance instructions are discussed in the NHANES Laboratory/Medical Technologists Procedures Manual (LPM). Read the LABDOC file for detailed QA/QC protocols.

Mobile Examination Centers (MECs)

Laboratory team performance is monitored using several techniques. NCHS and contract consultants use a structured quality assurance evaluation during unscheduled visits to evaluate both the quality of the laboratory work and the quality-control procedures. Each laboratory staff person is observed for equipment operation, specimen collection and preparation; testing procedures and constructive feedback are given to each staff. Formal retraining sessions are conducted annually to ensure that required skill levels are maintained.

Analytical Laboratories

NHANES uses several methods to monitor the quality of the analyses performed by the contract laboratories. In the MEC, these methods include performing blind split samples collected on “dry run” sessions. In addition, contract laboratories randomly perform repeat testing on 2.0% of all specimens.

NCHS developed and distributed a quality control protocol for all the contract laboratories which outlined the Westgard rules used when running NHANES specimens. Progress reports containing any problems encountered during shipping or receipt of specimens, summary statistics for each control pool, QC graphs, instrument calibration, reagents, and any special considerations are submitted to NCHS and Westat quarterly. The reports are reviewed for trends or shifts in the data. The laboratories are required to explain any identified areas of concern.

There were no changes to the site or laboratory from the previous two year cycle. Beginning in 2005, urinary iodine and mercury were tested from the same instrument.

All QC procedures recommended by the manufacturers were followed. Reported results for all assays meet the Division of Laboratory Science’s quality control and quality assurance performance criteria for accuracy and precision (similar to specifications outlined by Westgard, 1981).

Analytic Notes

Subsample weights

Measures of urinary mercury were conducted in a one third subsample of persons 6 years and over, therefore special sample weights are required to analyze these data properly. Specific sample weights for this subsample are included in this data file and should be used when analyzing these data.

Variance estimation

The analysis of NHANES laboratory data must be conducted with the key survey design and basic demographic variables. The NHANES Demographic Data File contains demographic and sample design variables. The recommended procedure for variance estimation requires use of stratum and PSU variables (SDMVSTRA and SDMVPSU, respectively) in the demographic data file.

Links to NHANES Data Files

This laboratory data file can be linked to the other NHANES 2007-2008 data files using the unique survey participant identifier SEQN.

Detection Limits

The lower detection limit for urinary mercury was constant during this two-year cycle.

Please refer to the Analytic Guidelines for further details on the use of sample weights and other analytic issues. The Analytic Guidelines are available on the NHANES website.

Codebook and Frequencies

SEQN - Respondent sequence number

Variable Name:
SEQN
SAS Label:
Respondent sequence number
English Text:
Respondent sequence number.
Target:
Both males and females 6 YEARS - 150 YEARS

URXUHG - Mercury, urine (ug/L)

Variable Name:
URXUHG
SAS Label:
Mercury, urine (ug/L)
English Text:
Mercury, urine (ug/L)
Target:
Both males and females 6 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0.06 to 34.42 Range of Values 2634 2634
. Missing 86 2720

URXUCR - Creatinine, urine (mg/dL)

Variable Name:
URXUCR
SAS Label:
Creatinine, urine (mg/dL)
English Text:
Creatinine, urine (mg/dL)
Target:
Both males and females 6 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
7 to 528 Range of Values 2657 2657
. Missing 63 2720

WTSA2YR - Environmental A 2 year weights

Variable Name:
WTSA2YR
SAS Label:
Environmental A 2 year weights
English Text:
Environmental A 2 year weights
Target:
Both males and females 6 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
0 to 604864.71127 Range of Values 2720 2720
. Missing 0 2720