Trace metals have been associated with adverse health effects in occupational studies or laboratory studies, but have not been monitored in general population groups.
This method is used to achieve rapid and accurate quantifications of three elements of toxicological and nutritional interest. The method is sensitive and rapid enough to screen serum specimens from subjects suspected to be exposed to a number of important toxic elements, or to evaluate environmental or other nonoccupational exposure to these same elements.
Examined participants aged 6 years and older from a one-third sample were eligible.
Inductively coupled plasma dynamic reaction cell mass spectrometry (ICP-DRC-MS) is a multi-element analytical technique capable of trace level elemental analysis. This ICP-DRC-MS method is used to measure the entire panel of the 3 elements, or any subgroup of these. Liquid samples are introduced into the ICP through a nebulizer and spray chamber carried by a flowing argon stream. By coupling radio-frequency power into flowing argon, plasma is created and the sample passes through a region of the plasma, the thermal energy atomizes the sample and then ionizes the atoms. The ions, along with the argon, enter the mass spectrometer through an interface that separates the ICP from the mass spectrometer. The ions pass through a focusing region, the dynamic reaction cell (DRC), the quadrupole mass filter, and finally are counted in rapid sequence at the detector allowing individual isotopes of an element to be determined. The isotopes measured by this method include zinc (m/z 64), copper (m/z 65), and selenium (m/z 78) and the internal standard gallium (m/z 71). Serum samples are diluted 1+1+28 with water and diluent containing gallium (Ga) for multi-internal standardization.
Refer to the Laboratory Method Files section for detailed laboratory procedure manual(s) of the methods used.
There were no changes to lab equipment, lab methods, or lab site in the 2013-2014 cycle.
Zinc, Copper and Selenium (January 2016)
Serum specimens are processed, stored, and shipped to the Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA for analysis.
Detailed instructions on specimen collection and processing are discussed in the NHANES Laboratory Procedures Manual (LPM). Vials are stored under appropriate frozen (–70°C) conditions until they are shipped to the National Center for Environmental Health for testing.
The NHANES quality assurance and quality control (QA/QC) protocols meet the 1988 Clinical Laboratory Improvement Act mandates. Detailed QA/QC instructions are discussed in the NHANES LPM.
Mobile Examination Centers (MECs)
Laboratory team performance is monitored using several techniques. NCHS and contract consultants use a structured quality assurance evaluation during unscheduled visits to evaluate both the quality of the laboratory work and the quality-control procedures. Each laboratory staff member is observed for equipment operation, specimen collection and preparation; testing procedures and constructive feedback are given to each staff member. Formal retraining sessions are conducted annually to ensure that required skill levels were maintained.
Analytical Laboratories
NHANES uses several methods to monitor the quality of the analyses performed by the contract laboratories. In the MEC, these methods include performing blind split samples collected on “dry run” sessions. In addition, contract laboratories randomly perform repeat testing on 2% of all specimens.
NCHS developed and distributed a quality control protocol for all the contract laboratories which outlined the use of Westgard rules (Westgard et al, 1981) when running NHANES specimens. Progress reports containing any problems encountered during shipping or receipt of specimens, summary statistics for each control pool, QC graphs, instrument calibration, reagents, and any special considerations are submitted to NCHS quarterly. The reports are reviewed for trends or shifts in the data. The laboratories are required to explain any identified areas of concern.
All QC procedures recommended by the manufacturers were followed. Reported results for all assays meet the Division of Laboratory Sciences’ quality control and quality assurance performance criteria for accuracy and precision, similar to the Westgard rules (Caudill et al, 2008).
The data were reviewed. Incomplete data or improbable values were sent to the performing laboratory for confirmation.
Refer to the 2013-2014 Laboratory Data Overview for general information on NHANES laboratory data.
Subsample Weights
Serum copper, selenium, and zinc were measured in a one-third subsample of participants aged 6 years and older. Special sample weights are required to analyze these data properly. Specific sample weights for this subsample are included in this data file and should be used when analyzing these data.
Demographic and Other Related Variables
The analysis of NHANES laboratory data must be conducted using the appropriate survey design and demographic variables. The NHANES 2013-2014 Demographics File contains demographic data, health indicators, and other related information collected during household interviews as well as the sample design variables. The recommended procedure for variance estimation requires use of stratum and PSU variables (SDMVSTRA and SDMVPSU, respectively) in the demographic data file.
The Fasting Questionnaire File includes auxiliary information such as fasting status, the time of venipuncture, and the conditions precluding venipuncture.
This laboratory data file can be linked to the other NHANES data files using the unique survey participant identifier (i.e., SEQN).
Detection Limits
The detection limits were constant for all of the analytes in the data set. The variable prefixed LBX (ex., LBXSCU) provides the analytic result for the analyte.
The lower limit of detection (LLOD, in µg/dL) for copper, zinc, and for selenium (LLOD, in ug/L) is:
Variable Name |
SAS Label |
LLOD |
LBXSCU |
Serum Copper |
2.5 |
LBXSSE |
Serum Selenium |
4.5 |
LBXSZN |
Serum Zinc |
2.9 |
Please refer to the NHANES Analytic Guidelines and the on-line NHANES Tutorial for further details on the use of sample weights and other analytic issues.
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
16284.37488 to 530325.34726 | Range of Values | 2724 | 2724 | |
0 | No lab result | 31 | 2755 | |
. | Missing | 0 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
24.7 to 297.5 | Range of Values | 2520 | 2520 | |
. | Missing | 235 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
3.88 to 46.71 | Range of Values | 2520 | 2520 | |
. | Missing | 235 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
74.9 to 297.9 | Range of Values | 2519 | 2519 | |
. | Missing | 236 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
0.95 to 3.78 | Range of Values | 2519 | 2519 | |
. | Missing | 236 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
38.9 to 160.9 | Range of Values | 2519 | 2519 | |
. | Missing | 236 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
5.95 to 24.62 | Range of Values | 2519 | 2519 | |
. | Missing | 236 | 2755 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
5 to 546 | Range of Values | 2681 | 2681 | |
. | Missing | 74 | 2755 |