The NHANES 2017-2018 audiometry examination component (variable name prefix AUX) consists of four parts:

1) Pre-exam audiometric question – a series of questions to identify conditions that would affect how audiometric testing is conducted, or how results are interpreted. Questions include whether the participant has ear tubes, a current cold or ear problem, recent loud noise exposure, and listened to music with earphones in the past 24 hours. NOTE: more extensive data relating to a survey participant’s hearing status and history of noise exposure are contained in the audiometry questionnaire (AUQ) section of the NHANES Sample Person Questionnaire.

2) Otoscopy – a brief otoscopic screening (physical) exam of the ear canals and eardrums to identify abnormalities, which would require alternate audiometric procedures or influence interpretation of test results, and to identify conditions which might require medical referral. The exam screened for excessive or impacted ear cerumen (wax), physical abnormalities, or collapsing external ear canals.

3) Middle ear testing – a measurement of eardrum compliance by placing a probe with a rubber tip into the entrance of the ear canal. It is used to identify middle ear pathologies that might contribute to a hearing loss. There are three middle ear tests: tympanometry, wideband reflectance, and acoustic reflex. The AUX_J data file consists of tympanometric quantitative measurement variables for middle ear volume, pressure, compliance, and gradient. It also includes tympanogram quality ratings and scores variables. The raw data for the participant's tympanometric curves (AUXTYM_J), wideband reflectance curves (AUXWBR_J), and acoustic reflex curves (AUXAR_J) are released separately.

4) Pure tone air conduction audiometry – a measurement of hearing sensitivity by presenting pure tone signals to the ear through earphones and by varying the intensity of the signals until a participant’s hearing threshold at that frequency is determined. Testing is performed at frequencies across the range of human hearing.

Eligible Sample

All participants aged 6–19 years and 70 years and older were eligible. There were no precluding conditions for any part of the audiologic exam. However, participants using hearing aids who were not able to remove them for testing and participants who had sufficient ear pain at the time of the exam that they could not tolerate headphones were excluded.

Protocol and Procedure

All audiometry exam sections were performed by trained examiners on participants in a dedicated sound-isolating room in the Mobile Examination Center (MEC). Hearing threshold testing was conducted on both ears of participants at seven frequencies. Children aged 6–11 years were tested in the following frequency order so that data may be obtained for more important frequencies first in case the child’s attention wanes before all frequencies can be completed: Ear 1: 1000, 2000, 1000 (repeat), 4000, 8000 Hz. Ear 2: 1000, 2000, 1000 (repeat), 4000, 8000, 6000, 3000, 500 Hz. Ear 1: 6000, 3000, 500 Hz. Adolescents aged 12–19 years and adults aged 70 years and older were tested in the frequency order 1000, 500, 1000 (repeat), 2000, 3000, 4000, 6000, 8000 Hz for ear 1 and then for ear 2. The ear tested first was varied to prevent biasing the data. However, if the participant indicated one ear hearing better than the other, the ear with better sensitivity was tested first. Testing was conducted according to a modified Hughson Westlake procedure using the automated testing mode of the audiometer. The approximate audiometer output constraints for standard and insert earphones was 100 decibels (dB) at 500 to 6000 Hz and 90 dB at 8000 Hz. Some earphones might produce slightly higher output than others. A threshold greater than 90 dB can be obtained using a manual testing mode. Observed values, therefore, varied between -10 and 110 dB. If a participant did not respond to the signal tone at any level for one or more frequencies because of deafness or severe hearing loss, a code value of 666 was entered. Manual testing was conducted on children aged 6-8 years. Manual testing was also conducted when the participant could not operate the response switch or responded too slowly for the audiometer to accurately record the response.

In some instances, if a pure tone audiometric signal is sufficiently loud, it can "crossover" and be heard by the opposite ear via bone conduction. When this occurs, it is difficult to determine if the threshold obtained is truly the threshold of the test ear, or an artifact of the non-test ear (which may have better hearing). For the NHANES hearing threshold testing, a crossover retesting protocol was performed whenever the observed threshold at any given frequency was poorer in one ear than the other by 25 dB at 500 and 1000 Hz or 40 dB at any higher frequency. Retesting was accomplished using insert earphones, which are smaller and have less direct contact with the bones of the head. Thus, a much louder stimulus is required before crossover occurs. Due to the complexity of the procedure, masking was not employed in the NHANES audiometry exam.

Quality Assurance & Quality Control

Instrumentation for the audiometry component included an Audiometric Research Tool (ART) system with standard TDH-49P headphones and Etymotic EarTone 3A insert earphones. Interacoustics Titan (middle ear analyzer) was used to conduct middle ear testing. Daily monitoring ambient noise levels was accomplished with a Quest Model BA-202-27 Bioacoustic Simulator. AUX data were entered directly into the computerized NHANES database system. Data from the ART system and Interacoustics Titan were captured electronically and uploaded into the survey information system automatically.

Beginning in the 2017-2018, the ART system was used to conduct the pure-tone air conduction hearing test on survey participants. The ART system is a comprehensive, high configurable audiometer designed for the research application. The ART system consists of a dedicated National Instruments hardware and ART software as an audiometer. The ART software has been programmed to conduct the threshold test automatically and can also be used to conduct the threshold test manually. The ART system also contains an integrated calibration system, called Trident, to measure the sound levels put out by the system. A G.R.A.S. calibration kit with couplers for each type of earphones is used in conjunction with the ART to monitor calibration. The ART calibrations include the Trident software itself check, the background noise measurement, an acoustic calibration check (to measure the output, distortion, and linearity of the audiometer through both standard and insert earphones), and a listening check (tone quality, attenuator accuracy, headphone cords, crossover, etc.). The ART calibration checks were conducted at the beginning of each survey location, daily throughout the entire exam, and at the end of each survey location to ensure that the accuracy has not shifted. In addition to these calibration checks, each ART system received an exhaustive calibration once a year or whenever unresolvable problems were discovered during the calibration checks. The calibration must be accomplished by the manufacturer or a certified laboratory to ensure that the calibrations are traceable to the National Institute of Standards and Technology (NIST).

An Interacoustics Titan was used to conduct middle ear testing. The physical volume was checked on the Titan to ensure that it was within the calibration limits. In addition, the MEC Health Technicians ran a sample test on one ear to ensure that the unit is producing clear, normal graphs for the tympanogram and wideband reflectance. The Titan calibration checks were conducted at the beginning of each survey location, daily throughout the entire exam, and at the end each survey location. The Titan also received an exhaustive NIST-traceable calibration check annually.

A Bioacoustic Simulator was used to continuously measure the background noise levels in the audiometric test room during the exam. Pure tone audiometric testing was not performed if ambient noise levels in the test booth exceeded maximum permissible levels. The microphones and sound measurement function of the bioacoustic simulator were checked by a certified laboratory annually or whenever unsolvable problems were discovered.

As an additional quality measure, all audiograms, whether conducted in automated or manual mode, tested the 1000 Hz frequency twice in each ear as a measure of the reliability of the participant’s responses. Pure tone audiograms were not accepted if there was more than a 10 dB difference between them. For further details regarding any of these procedures, analysts should consult the NHANES 2017-2018 Audiometry Procedures Manual.

The MEC health technicians who performed the audiometry examination component of NHANES were professionally trained by a certified audiologist from the National Institute for Occupational Safety & Health (NIOSH). NIOSH also monitored the performance of each health technician on a regular basis. Field visits to each MEC were conducted by the NIOSH audiologist at least three times per year. Additionally, NCHS project officers visited the MECs approximately twice per year to observe the audiometry examinations and verify that standard testing procedures were being followed. NIOSH consultants provided the MEC health technicians with annual retraining and protocol updates.

The quantitative tympanometry measurements reported in this data file represent the single best of each of the two curves obtained during middle ear testing. Subjective quality ratings for these tympanograms are the variables AUAREQC and AUALEQC. These represent a qualitative assessment of each tympanogram by the consulting audiologist. The consulting audiologist also scored these tympanograms using a common classification system developed by Liden (1969) and modified by Jerger (1970). Two variables AUATYMTR and AUATYMTL provide information about what the tympanometry results mean in terms of the participant’s middle ear status.

Data Processing and Editing

The wording of some questions asked before the audiometry exam has been simplified for children aged 6–11 years. Therefore, there are two separated sets of variables for the pre-exam audiometric questionnaire: one for children aged 6–11 years and another for adolescents aged 12–19 years and adults aged 70 years and older.

All data were captured in the NHANES computerized database system, with audiometric and middle ear analyzer data automatically uploaded. On a continuous basis, a consulting audiologist performed a clinical review of all data for each participant as it was received, checking for quality and consistency. In addition, a computerized data editing program was developed to check for logical inconsistencies in the data and technician errors, and to cross-check other issues affecting data quality (consistency in identifying potential instances where crossover effects might have occurred, assurance of randomization of the initial test ear, etc.). Back-end edits of the data were performed as needed when errors were detected.

Analytic Notes

This examination was administered to a full sample of participants in the MEC. Therefore, the NHANES MEC examination weights should be used for data analysis. For more information on the use of sample weights in current NHANES data analysis and for combining data across multiple NHANES survey years, please refer to the NHANES Analytic Guidelines and the online NHANES Tutorial for further details on the use of sample weights and other analytic issues.

Audiologic data analysis is a complex procedure and requires a thorough knowledge of the specialty content area for valid results to be obtained. If an analyst does not have professional experience in this area, it is recommended that audiologic consultation be obtained to help formulate and review the results of the analysis. Data analysts should be especially aware of the fact that the number "666" in all primary audiometric frequency data fields (AUXU1K1R through AUXU8KL) as well as in all repeat test frequencies (AUXR1K1R through AUXR8KL) is a qualitative code for nonresponse at a particular frequency and does not represent actual measured decibel hearing threshold values. These nonresponse values should be appropriately edited prior to any numerical data analysis.

References

Codebook and Frequencies

SEQN - Respondent sequence number

Variable Name:
SEQN
SAS Label:
Respondent sequence number
English Text:
Respondent sequence number.
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS

AUAEXSTS - Status of an Audio Exam

Variable Name:
AUAEXSTS
SAS Label:
Status of an Audio Exam
English Text:
Status of an Audiometry Exam
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Complete 2343 2343
2 Partial 609 2952
3 Not done 179 3131
. Missing 0 3131

AUAEXCMT - Comment code for an Audio Exam

Variable Name:
AUAEXCMT
SAS Label:
Comment code for an Audio Exam
English Text:
Comment code for an Audio Exam
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
2 SP refusal 45 45
3 No time 22 67
4 Physical limitation 161 228
5 Communication problem 5 233
6 Equipment failure 23 256
7 SP ill/emergency 5 261
14 Interrupted 1 262
51 SP unable to comply 319 581
56 Came late/left early 129 710
72 Error (technician/software/supply) 33 743
99 Other, specify 24 767
122 Language barrier 2 769
150 SP with parent SP 2 771
. Missing 2360 3131

AUQ011 - Have Ear Tube, Right or Left Ear?

Variable Name:
AUQ011
SAS Label:
Have Ear Tube, Right or Left Ear?
English Text:
Do you now have a tube in your right or left ear? (If yes, indicate affected ear(s))
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes, right ear 2 2
2 Yes, left ear 5 7
3 Yes, both ears 14 21
4 No 948 969
7 Refused 4 973
9 Don't know 5 978
. Missing 2153 3131

AUQ540 - Had Cold, Runny Nose, Earache Today?

Variable Name:
AUQ540
SAS Label:
Had Cold, Runny Nose, Earache Today?
English Text:
Do you have a cold, runny nose, or earache today?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 170 170
2 No 800 970
7 Refused 0 970
9 Don't know 5 975
. Missing 2156 3131

AUQ550 - Had Cold, Runny Nose, Earache Yesterday?

Variable Name:
AUQ550
SAS Label:
Had Cold, Runny Nose, Earache Yesterday?
English Text:
Did you have a cold, runny nose, or earache yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 56 56
2 No 741 797
7 Refused 1 798
9 Don't know 5 803
. Missing 2328 3131

AUQ550A - Had Cold?

Variable Name:
AUQ550A
SAS Label:
Had Cold?
English Text:
Have you had a cold today or yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 42 42
2 No (checkbox unchecked) 184 226
9 Don't know 0 226
. Missing 2905 3131

AUQ550B - Had Runny Nose?

Variable Name:
AUQ550B
SAS Label:
Had Runny Nose?
English Text:
Have you had runny nose today or yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 182 182
2 No (checkbox unchecked) 44 226
9 Don't know 0 226
. Missing 2905 3131

AUQ550C - Earache, Right Ear?

Variable Name:
AUQ550C
SAS Label:
Earache, Right Ear?
English Text:
Have you had an earache in right ear today or yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 6 6
2 No (checkbox unchecked) 220 226
9 Don't know 0 226
. Missing 2905 3131

AUQ550D - Earache, Left Ear?

Variable Name:
AUQ550D
SAS Label:
Earache, Left Ear?
English Text:
Have you had an earache in left ear today or yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 8 8
2 No (checkbox unchecked) 218 226
9 Don't know 0 226
. Missing 2905 3131

AUQ550E - Earache, Both Ears?

Variable Name:
AUQ550E
SAS Label:
Earache, Both Ears?
English Text:
Have you had an earache in both ears today or yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 2 2
2 No (checkbox unchecked) 224 226
9 Don't know 0 226
. Missing 2905 3131

AUQ560 - Listened to Loud Sound Today?

Variable Name:
AUQ560
SAS Label:
Listened to Loud Sound Today?
English Text:
Have you listened to sounds today that were so loud you would have to shout so someone close by could hear you?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 56 56
2 No 906 962
7 Refused 2 964
9 Don't know 8 972
. Missing 2159 3131

AUQ570 - Listened to Loud Sound Yesterday?

Variable Name:
AUQ570
SAS Label:
Listened to Loud Sound Yesterday?
English Text:
Did you listen to any sounds that were that loud yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 50 50
2 No 857 907
7 Refused 0 907
9 Don't know 12 919
. Missing 2212 3131

AUQ580 - Hours Since Loud Sound Ended?

Variable Name:
AUQ580
SAS Label:
Hours Since Loud Sound Ended?
English Text:
When was that?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 to 23 Range of Values 96 96
77 Refused 0 96
99 Don't know 10 106
. Missing 3025 3131

AUQ590 - Listened to Music with Earphones Today?

Variable Name:
AUQ590
SAS Label:
Listened to Music with Earphones Today?
English Text:
Have you listened to music with earphones today?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 100 100
2 No 874 974
7 Refused 0 974
9 Don't know 2 976
. Missing 2155 3131

AUQ600 - Listened to Music w/Earphones Yesterday?

Variable Name:
AUQ600
SAS Label:
Listened to Music w/Earphones Yesterday?
English Text:
Did you listen to music with earphones yesterday?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 91 91
2 No 780 871
7 Refused 0 871
9 Don't know 4 875
. Missing 2256 3131

AUQ610 - Hours Since Stopped Listening?

Variable Name:
AUQ610
SAS Label:
Hours Since Stopped Listening?
English Text:
When was that?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 to 24 Range of Values 177 177
77 Refused 0 177
99 Don't know 14 191
. Missing 2940 3131

AUQ051 - Self Reported Better Ear

Variable Name:
AUQ051
SAS Label:
Self Reported Better Ear
English Text:
Is it easier for you to hear out of one ear than the other?
Target:
Both males and females 6 YEARS - 11 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes, right ear 59 59
2 Yes, left ear 41 100
7 Refused 0 100
9 No / don't know 876 976
. Missing 2155 3131

AUQ010 - Have Ear Tube, Right or Left Ear?

Variable Name:
AUQ010
SAS Label:
Have Ear Tube, Right or Left Ear?
English Text:
Do you now have a tube in your right or left ear? (If yes, indicate affected ear(s))
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes, right ear 4 4
2 Yes, left ear 2 6
3 Yes, both ears 1 7
4 No 1952 1959
7 Refused 1 1960
9 Don't know 14 1974
. Missing 1157 3131

AUQ020 - Had Cold, Sinus or Earache Last 24 Hrs?

Variable Name:
AUQ020
SAS Label:
Had Cold, Sinus or Earache Last 24 Hrs?
English Text:
Have you had a cold, sinus problem, or earache in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 184 184
2 No 1785 1969
7 Refused 2 1971
9 Don't know 3 1974
. Missing 1157 3131

AUQ020A - Had Cold Last 24 Hrs?

Variable Name:
AUQ020A
SAS Label:
Had Cold Last 24 Hrs?
English Text:
Have you had a cold in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 59 59
2 No (checkbox unchecked) 125 184
9 Don't know 0 184
. Missing 2947 3131

AUQ020B - Had Sinus Problem Last 24 Hours?

Variable Name:
AUQ020B
SAS Label:
Had Sinus Problem Last 24 Hours?
English Text:
Have you had a sinus problem in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 121 121
2 No (checkbox unchecked) 63 184
9 Don't know 0 184
. Missing 2947 3131

AUQ020C - Earache Last 24 Hours, Right?

Variable Name:
AUQ020C
SAS Label:
Earache Last 24 Hours, Right?
English Text:
Have you had an earache in the right ear in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 6 6
2 No (checkbox unchecked) 178 184
9 Don't know 0 184
. Missing 2947 3131

AUQ020D - Earache Last 24 Hours, Left?

Variable Name:
AUQ020D
SAS Label:
Earache Last 24 Hours, Left?
English Text:
Have you had an earache in left ear in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 9 9
2 No (checkbox unchecked) 175 184
9 Don't know 0 184
. Missing 2947 3131

AUQ020E - Earache Last 24 Hours, Both?

Variable Name:
AUQ020E
SAS Label:
Earache Last 24 Hours, Both?
English Text:
Have you had an earache in both ears in the last 24 hours?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 2 2
2 No (checkbox unchecked) 182 184
9 Don't know 0 184
. Missing 2947 3131

AUQ031 - Exposed to Loud Noise Last 24 Hrs?

Variable Name:
AUQ031
SAS Label:
Exposed to Loud Noise Last 24 Hrs?
English Text:
In the past 24 hours, have you been exposed to noise so loud that you would have had to raise your voice to speak to someone an arm's length away?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 144 144
2 No 1829 1973
7 Refused 0 1973
9 Don't know 1 1974
. Missing 1157 3131

AUQ041 - Hours Since Noise Ended?

Variable Name:
AUQ041
SAS Label:
Hours Since Noise Ended?
English Text:
How many hours ago did the noise end?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
1 to 24
Code or Value Value Description Count Cumulative Skip to Item
1 to 24 Range of Values 143 143
77 SP refused 0 143
99 Don't know 1 144
. Missing 2987 3131

AUQ520 - Listened to Music w/Earphone Last 24 Hrs

Variable Name:
AUQ520
SAS Label:
Listened to Music w/Earphone Last 24 Hrs
English Text:
In the past 24 hours, have you listened to music with earphones?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes 560 560
2 No 1412 1972
7 Refused 0 1972
9 Don't know 2 1974
. Missing 1157 3131

AUQ530 - Hours Since Music Ended?

Variable Name:
AUQ530
SAS Label:
Hours Since Music Ended?
English Text:
How many hours ago did you stop listening?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 to 24 Range of Values 555 555
77 Refused 0 555
99 Don't know 5 560
. Missing 2571 3131

AUQ050 - Self Reported Better Ear

Variable Name:
AUQ050
SAS Label:
Self Reported Better Ear
English Text:
Do you hear better in one ear than the other?
Target:
Both males and females 12 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes, right ear 218 218
2 Yes, left ear 187 405
7 Refused 1 406
9 No / don't know 1568 1974
. Missing 1157 3131

AUXOTSPL - Normal Otoscopy, Left Ear?

Variable Name:
AUXOTSPL
SAS Label:
Normal Otoscopy, Left Ear?
English Text:
Normal: Left Ear Otoscopy
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 2382 2382
2 No (checkbox unchecked) 568 2950
. Missing 181 3131

AUXLOEXC - Excessive Cerumen, Left Ear

Variable Name:
AUXLOEXC
SAS Label:
Excessive Cerumen, Left Ear
English Text:
Excessive Cerumen, Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 532 532
2 No (checkbox unchecked) 2418 2950
. Missing 181 3131

AUXLOIMC - Impacted Cerumen, Left Ear

Variable Name:
AUXLOIMC
SAS Label:
Impacted Cerumen, Left Ear
English Text:
Impacted Cerumen, Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 29 29
2 No (checkbox unchecked) 2921 2950
. Missing 181 3131

AUXLOCOL - Collapsing Ear Canals, Left Ear

Variable Name:
AUXLOCOL
SAS Label:
Collapsing Ear Canals, Left Ear
English Text:
Collapsing Ear Canal, Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 3 3
2 No (checkbox unchecked) 2947 2950
. Missing 181 3131

AUXLOABN - Other Exam Abnormality, Left Ear?

Variable Name:
AUXLOABN
SAS Label:
Other Exam Abnormality, Left Ear?
English Text:
Other Exam Abnormality, Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 6 6
2 No (checkbox unchecked) 2944 2950
. Missing 181 3131

AUDLOABC - Comment,Other Exam Abnormality Left Ear

Variable Name:
AUDLOABC
SAS Label:
Comment,Other Exam Abnormality Left Ear
English Text:
Other Exam Abnormality, Left, Comment
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Possible Perforated Eardrum 3 3
3 Possible Infection 0 3
4 TM Scarring or Deformity 0 3
5 Possible Ear Canal Abrasion 1 4
6 Growth or Foreign Body in Ear Canal 1 5
7 Deformity of Auricle 0 5
8 Other 1 6
9 Could not Visualize Eardrum 0 6
. Missing 3125 3131

AUXROTSP - Normal Otoscopy, Right Ear?

Variable Name:
AUXROTSP
SAS Label:
Normal Otoscopy, Right Ear?
English Text:
Normal: Right Ear Otoscopy
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 2371 2371
2 No (checkbox unchecked) 579 2950
. Missing 181 3131

AUXROEXC - Excessive Cerumen, Right Ear

Variable Name:
AUXROEXC
SAS Label:
Excessive Cerumen, Right Ear
English Text:
Excessive Cerumen, Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 543 543
2 No (checkbox unchecked) 2407 2950
. Missing 181 3131

AUXROIMC - Impacted Cerumen, Right Ear

Variable Name:
AUXROIMC
SAS Label:
Impacted Cerumen, Right Ear
English Text:
Impacted Cerumen, Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 30 30
2 No (checkbox unchecked) 2920 2950
. Missing 181 3131

AUXROCOL - Collapsing Ear Canals, Right Ear

Variable Name:
AUXROCOL
SAS Label:
Collapsing Ear Canals, Right Ear
English Text:
Collapsing Ear Canal, Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 4 4
2 No (checkbox unchecked) 2946 2950
. Missing 181 3131

AUXROABN - Other Exam Abnormality, Right Ear?

Variable Name:
AUXROABN
SAS Label:
Other Exam Abnormality, Right Ear?
English Text:
Other Exam Abnormality, Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Yes (checkbox checked) 6 6
2 No (checkbox unchecked) 2944 2950
. Missing 181 3131

AUDROABC - Comment,Other Exam Abnormality Right Ear

Variable Name:
AUDROABC
SAS Label:
Comment,Other Exam Abnormality Right Ear
English Text:
Other Exam Abnormality, Right, Comment
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Possible Perforated Eardrum 2 2
3 Possible Infection 0 2
4 TM Scarring or Deformity 0 2
5 Possible Ear Canal Abrasion 0 2
6 Growth or Foreign Body in Ear Canal 1 3
7 Deformity of Auricle 0 3
8 Other 2 5
9 Could not Visualize Eardrum 1 6
. Missing 3125 3131

AUXTMEPR - Middle Ear Pressure-Tymp, Rt Ear in daPa

Variable Name:
AUXTMEPR
SAS Label:
Middle Ear Pressure-Tymp, Rt Ear in daPa
English Text:
Middle ear pressure (tympanometry), right ear, in daPa (dekaPascals)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
-400.0000 to 200.0000
Code or Value Value Description Count Cumulative Skip to Item
-299 to 199 Range of Values 2654 2654
. Missing 477 3131

AUXTPVR - Physical Volume-Tymp, Right Ear in cc

Variable Name:
AUXTPVR
SAS Label:
Physical Volume-Tymp, Right Ear in cc
English Text:
Physical volume (tympanometry), right ear, in cc
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0 to 7.0
Code or Value Value Description Count Cumulative Skip to Item
0.214 to 6.478 Range of Values 2782 2782
. Missing 349 3131

AUXTWIDR - Tympanometric width, right ear

Variable Name:
AUXTWIDR
SAS Label:
Tympanometric width, right ear
English Text:
Tympanometric width, right ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0000 to 500.0000
Code or Value Value Description Count Cumulative Skip to Item
1 to 444 Range of Values 2655 2655
. Missing 476 3131

AUXTCOMR - Compliance (tympanometry), right ear

Variable Name:
AUXTCOMR
SAS Label:
Compliance (tympanometry), right ear
English Text:
Compliance (tympanometry), right ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0000 to 8.0000
Code or Value Value Description Count Cumulative Skip to Item
0.15 to 6.233 Range of Values 2656 2656
. Missing 475 3131

AUXTMEPL - Middle Ear Pressure, Tymp, Left in daPa

Variable Name:
AUXTMEPL
SAS Label:
Middle Ear Pressure, Tymp, Left in daPa
English Text:
Middle ear pressure (tympanometry), left ear, in daPa (dekaPascals)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
-400.0000 to 200.0000
Code or Value Value Description Count Cumulative Skip to Item
-299 to 173 Range of Values 2661 2661
. Missing 470 3131

AUXTPVL - Physical Volume-Tymp, Left ear in cc

Variable Name:
AUXTPVL
SAS Label:
Physical Volume-Tymp, Left ear in cc
English Text:
Physical volume (tympanometry), left ear, in cc
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0 to 7.0
Code or Value Value Description Count Cumulative Skip to Item
0.09 to 5.234 Range of Values 2778 2778
. Missing 353 3131

AUXTWIDL - Tympanometric width, left ear

Variable Name:
AUXTWIDL
SAS Label:
Tympanometric width, left ear
English Text:
Tympanometric width, left ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0000 to 500.0000
Code or Value Value Description Count Cumulative Skip to Item
17 to 435 Range of Values 2662 2662
. Missing 469 3131

AUXTCOML - Compliance (tympanometry), left ear

Variable Name:
AUXTCOML
SAS Label:
Compliance (tympanometry), left ear
English Text:
Compliance (tympanometry), left ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Hard Edits:
0.0000 to 8.0000
Code or Value Value Description Count Cumulative Skip to Item
0.15 to 5.793 Range of Values 2664 2664
. Missing 467 3131

AUAEAR - Which ear tested first?

Variable Name:
AUAEAR
SAS Label:
Which ear tested first?
English Text:
Which ear tested first?
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Right 1464 1464
2 Left 1462 2926
. Missing 205 3131

AUAMODE - Audio Test Mode-Manual/Automatic/Mixed

Variable Name:
AUAMODE
SAS Label:
Audio Test Mode-Manual/Automatic/Mixed
English Text:
Mode of Audiometric Test (Manual/Automatic/Mixed)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Manual 502 502
2 Automatic 2110 2612
3 Mixed 309 2921
. Missing 210 3131

AUAFMANL - Frequency, Switch to Manual Mode, Left

Variable Name:
AUAFMANL
SAS Label:
Frequency, Switch to Manual Mode, Left
English Text:
Frequency at Which Switched From Auto to Manual Audio Testing (Left)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 None 2749 2749
2 500 Hz 6 2755
3 1000 Hz 25 2780
4 2000 Hz 17 2797
5 3000 Hz 5 2802
6 4000 Hz 12 2814
7 6000 Hz 25 2839
8 8000 Hz 82 2921
. Missing 210 3131

AUAFMANR - Frequency, Switch to Manual Mode, Right

Variable Name:
AUAFMANR
SAS Label:
Frequency, Switch to Manual Mode, Right
English Text:
Frequency at Which Switched From Auto to Manual Audio Testing (Right)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 None 2770 2770
2 500 Hz 9 2779
3 1000 Hz 24 2803
4 2000 Hz 10 2813
5 3000 Hz 15 2828
6 4000 Hz 10 2838
7 6000 Hz 23 2861
8 8000 Hz 60 2921
. Missing 210 3131

AUXU1K1R - Right threshold @ 1000Hz (db)

Variable Name:
AUXU1K1R
SAS Label:
Right threshold @ 1000Hz (db)
English Text:
Right threshold @ 1000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2781 2781
666 No response 0 2781
888 Could not obtain 101 2882
. Missing 249 3131

AUXU500R - Right threshold @ 500Hz (db)

Variable Name:
AUXU500R
SAS Label:
Right threshold @ 500Hz (db)
English Text:
Right threshold @ 500Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2613 2613
666 No response 1 2614
888 Could not obtain 264 2878
. Missing 253 3131

AUXU1K2R - Right Threshold @ 1000Hz-2nd Read (db)

Variable Name:
AUXU1K2R
SAS Label:
Right Threshold @ 1000Hz-2nd Read (db)
English Text:
Right threshold @ 1000Hz (second reading) in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2781 2781
666 No response 0 2781
888 Could not obtain 101 2882
. Missing 249 3131

AUXU2KR - Right threshold @ 2000Hz (db)

Variable Name:
AUXU2KR
SAS Label:
Right threshold @ 2000Hz (db)
English Text:
Right threshold @ 2000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 110 Range of Values 2773 2773
666 No response 4 2777
888 Could not obtain 104 2881
. Missing 250 3131

AUXU3KR - Right threshold @ 3000Hz (db)

Variable Name:
AUXU3KR
SAS Label:
Right threshold @ 3000Hz (db)
English Text:
Right threshold @ 3000Hz in decibels (Hearing Levels)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2585 2585
666 No response 10 2595
888 Could not obtain 280 2875
. Missing 256 3131

AUXU4KR - Right threshold @ 4000Hz (db)

Variable Name:
AUXU4KR
SAS Label:
Right threshold @ 4000Hz (db)
English Text:
Right threshold @ 4000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 110 Range of Values 2739 2739
666 No response 22 2761
888 Could not obtain 119 2880
. Missing 251 3131

AUXU6KR - Right threshold @ 6000Hz (db)

Variable Name:
AUXU6KR
SAS Label:
Right threshold @ 6000Hz (db)
English Text:
Right threshold @ 6000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 100 Range of Values 2539 2539
666 No response 50 2589
888 Could not obtain 283 2872
. Missing 259 3131

AUXU8KR - Right threshold @ 8000Hz (db)

Variable Name:
AUXU8KR
SAS Label:
Right threshold @ 8000Hz (db)
English Text:
Right threshold @ 8000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 95 Range of Values 2582 2582
666 No response 162 2744
888 Could not obtain 133 2877
. Missing 254 3131

AUXU1K1L - Left threshold @ 1000Hz (db)

Variable Name:
AUXU1K1L
SAS Label:
Left threshold @ 1000Hz (db)
English Text:
Left threshold @ 1000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 100 Range of Values 2770 2770
666 No response 2 2772
888 Could not obtain 108 2880
. Missing 251 3131

AUXU500L - Left threshold @ 500Hz (db)

Variable Name:
AUXU500L
SAS Label:
Left threshold @ 500Hz (db)
English Text:
Left threshold @ 500Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2608 2608
666 No response 2 2610
888 Could not obtain 267 2877
. Missing 254 3131

AUXU1K2L - Left threshold @ 1000Hz-2nd Read (db)

Variable Name:
AUXU1K2L
SAS Label:
Left threshold @ 1000Hz-2nd Read (db)
English Text:
Left threshold @ 1000Hz (second reading) in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 100 Range of Values 2770 2770
666 No response 2 2772
888 Could not obtain 108 2880
. Missing 251 3131

AUXU2KL - Left threshold @ 2000Hz (db)

Variable Name:
AUXU2KL
SAS Label:
Left threshold @ 2000Hz (db)
English Text:
Left threshold @ 2000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2760 2760
666 No response 5 2765
888 Could not obtain 114 2879
. Missing 252 3131

AUXU3KL - Left threshold @ 3000Hz (db)

Variable Name:
AUXU3KL
SAS Label:
Left threshold @ 3000Hz (db)
English Text:
Left threshold @ 3000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 110 Range of Values 2578 2578
666 No response 16 2594
888 Could not obtain 283 2877
. Missing 254 3131

AUXU4KL - Left threshold @ 4000Hz (db)

Variable Name:
AUXU4KL
SAS Label:
Left threshold @ 4000Hz (db)
English Text:
Left threshold @ 4000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 110 Range of Values 2730 2730
666 No response 25 2755
888 Could not obtain 124 2879
. Missing 252 3131

AUXU6KL - Left threshold @ 6000Hz (db)

Variable Name:
AUXU6KL
SAS Label:
Left threshold @ 6000Hz (db)
English Text:
Left threshold @ 6000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 105 Range of Values 2541 2541
666 No response 50 2591
888 Could not obtain 286 2877
. Missing 254 3131

AUXU8KL - Left threshold @ 8000Hz (db)

Variable Name:
AUXU8KL
SAS Label:
Left threshold @ 8000Hz (db)
English Text:
Left threshold @ 8000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
-10 to 95 Range of Values 2590 2590
666 No response 149 2739
888 Could not obtain 138 2877
. Missing 254 3131

AUXR1K1R - Right retest threshold @ 1000Hz (db)

Variable Name:
AUXR1K1R
SAS Label:
Right retest threshold @ 1000Hz (db)
English Text:
Right retest threshold @ 1000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
10 to 100 Range of Values 19 19
666 No response 0 19
888 Could not obtain 1 20
. Missing 3111 3131

AUXR5CR - Right retest threshold @ 500Hz (db)

Variable Name:
AUXR5CR
SAS Label:
Right retest threshold @ 500Hz (db)
English Text:
Right retest threshold @ 500Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
5 to 105 Range of Values 21 21
666 No response 0 21
888 Could not obtain 1 22
. Missing 3109 3131

AUXR1K2R - Right Retest Threshold 1000Hz-2nd Read

Variable Name:
AUXR1K2R
SAS Label:
Right Retest Threshold 1000Hz-2nd Read
English Text:
Right retest threshold @ 1000Hz, (second reading) in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
30 to 90 Range of Values 2 2
666 No response 0 2
888 Could not obtain 0 2
. Missing 3129 3131

AUXR2KR - Right retest threshold @ 2000Hz (db)

Variable Name:
AUXR2KR
SAS Label:
Right retest threshold @ 2000Hz (db)
English Text:
Right retest threshold @ 2000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
25 to 100 Range of Values 11 11
666 No response 0 11
888 Could not obtain 1 12
. Missing 3119 3131

AUXR3KR - Right retest threshold @ 3000Hz (db)

Variable Name:
AUXR3KR
SAS Label:
Right retest threshold @ 3000Hz (db)
English Text:
Right retest threshold @ 3000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
45 to 90 Range of Values 6 6
666 No response 1 7
888 Could not obtain 1 8
. Missing 3123 3131

AUXR4KR - Right retest threshold @ 4000Hz (db)

Variable Name:
AUXR4KR
SAS Label:
Right retest threshold @ 4000Hz (db)
English Text:
Right retest threshold @ 4000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
45 to 100 Range of Values 8 8
666 No response 1 9
888 Could not obtain 1 10
. Missing 3121 3131

AUXR6KR - Right retest threshold @ 6000Hz (db)

Variable Name:
AUXR6KR
SAS Label:
Right retest threshold @ 6000Hz (db)
English Text:
Right retest threshold @ 6000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
50 to 90 Range of Values 7 7
666 No response 2 9
888 Could not obtain 1 10
. Missing 3121 3131

AUXR8KR - Right retest threshold @ 8000Hz (db)

Variable Name:
AUXR8KR
SAS Label:
Right retest threshold @ 8000Hz (db)
English Text:
Right retest threshold @ 8000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
45 to 80 Range of Values 4 4
666 No response 1 5
888 Could not obtain 1 6
. Missing 3125 3131

AUXR1K1L - Left retest threshold @ 1000Hz (db)

Variable Name:
AUXR1K1L
SAS Label:
Left retest threshold @ 1000Hz (db)
English Text:
Left retest threshold @ 1000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
30 to 95 Range of Values 20 20
666 No response 0 20
888 Could not obtain 0 20
. Missing 3111 3131

AUXR5CL - Left retest threshold @ 500Hz (db)

Variable Name:
AUXR5CL
SAS Label:
Left retest threshold @ 500Hz (db)
English Text:
Left retest threshold @ 500Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
25 to 105 Range of Values 16 16
666 No response 0 16
888 Could not obtain 0 16
. Missing 3115 3131

AUXR1K2L - Left Retest Threshold 1000Hz-2nd Read

Variable Name:
AUXR1K2L
SAS Label:
Left Retest Threshold 1000Hz-2nd Read
English Text:
Left retest threshold @ 1000Hz, (second reading) in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
30 to 75 Range of Values 2 2
666 No response 0 2
888 Could not obtain 0 2
. Missing 3129 3131

AUXR2KL - Left retest threshold @ 2000Hz (db)

Variable Name:
AUXR2KL
SAS Label:
Left retest threshold @ 2000Hz (db)
English Text:
Left retest threshold @ 2000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
35 to 95 Range of Values 7 7
666 No response 2 9
888 Could not obtain 0 9
. Missing 3122 3131

AUXR3KL - Left retest threshold @ 3000Hz (db)

Variable Name:
AUXR3KL
SAS Label:
Left retest threshold @ 3000Hz (db)
English Text:
Left retest threshold @ 3000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
50 to 105 Range of Values 6 6
666 No response 0 6
888 Could not obtain 0 6
. Missing 3125 3131

AUXR4KL - Left retest threshold @ 4000Hz (db)

Variable Name:
AUXR4KL
SAS Label:
Left retest threshold @ 4000Hz (db)
English Text:
Left retest threshold @ 4000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
45 to 100 Range of Values 5 5
666 No response 0 5
888 Could not obtain 0 5
. Missing 3126 3131

AUXR6KL - Left retest threshold @ 6000Hz (db)

Variable Name:
AUXR6KL
SAS Label:
Left retest threshold @ 6000Hz (db)
English Text:
Left retest threshold @ 6000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
40 to 95 Range of Values 6 6
666 No response 1 7
888 Could not obtain 0 7
. Missing 3124 3131

AUXR8KL - Left retest threshold @ 8000Hz (db)

Variable Name:
AUXR8KL
SAS Label:
Left retest threshold @ 8000Hz (db)
English Text:
Left retest threshold @ 8000Hz in decibels (Hearing Level)
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
45 to 80 Range of Values 9 9
666 No response 1 10
888 Could not obtain 1 11
. Missing 3120 3131

AUAREQC - Right Ear Tympanogram Quality Code

Variable Name:
AUAREQC
SAS Label:
Right Ear Tympanogram Quality Code
English Text:
Quality Code for Tympanogram of Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Good - tympanogram is smooth and symmetrical 2298 2298
2 Adequate - tympanogram is 'noisy' but adequate for interpretation 438 2736
3 Poor - tympanogram cannot be interpreted 49 2785
7 SP Refused Tympanometry 0 2785
8 Could not obtain 116 2901
9 Equivocal - tympanogram cannot be evaluated without further information 12 2913
. Missing 218 3131

AUATYMTR - Tympanogram Type, Right Ear

Variable Name:
AUATYMTR
SAS Label:
Tympanogram Type, Right Ear
English Text:
Tympanogram Type, Right Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
Type A Normal 2487 2487
Type AS Normal, but the ear drum is unusually stiff 120 2607
Type AD Normal, but the ear drum is unusually flaccid 21 2628
Type B Flat, no measurable ear drum mobility 29 2657
Type C Normal ear drum mobility, but negative middle ear pressure 81 2738
88888 Could not obtain 116 2854
99999 Could not interpret 59 2913
< blank > Missing 218 3131

AUALEQC - Left Ear Tympanogram Quality Code

Variable Name:
AUALEQC
SAS Label:
Left Ear Tympanogram Quality Code
English Text:
Quality Code for Tympanogram of Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
1 Good - tympanogram is smooth and symmetrical 2326 2326
2 Adequate - tympanogram is 'noisy' but adequate for interpretation 402 2728
3 Poor - tympanogram cannot be interpreted 50 2778
7 SP Refused Tympanometry 0 2778
8 Could not obtain 125 2903
9 Equivocal - tympanogram cannot be evaluated without further information 10 2913
. Missing 218 3131

AUATYMTL - Tympanogram Type, Left Ear

Variable Name:
AUATYMTL
SAS Label:
Tympanogram Type, Left Ear
English Text:
Tympanogram Type, Left Ear
Target:
Both males and females 6 YEARS - 19 YEARS
Target:
Both males and females 70 YEARS - 150 YEARS
Code or Value Value Description Count Cumulative Skip to Item
Type A Normal 2499 2499
Type AS Normal, but the ear drum is unusually stiff 116 2615
Type AD Normal, but the ear drum is unusually flaccid 23 2638
Type B Flat, no measurable ear drum mobility 24 2662
Type C Normal ear drum mobility, but negative middle ear pressure 68 2730
88888 Could not obtain 125 2855
99999 Could not interpret 58 2913
< blank > Missing 218 3131