Heart disease is the leading cause of death in the United States (Murphy, et. al., 2018). Blood lipid levels are fundamental measures included in NHANES that can be used for cardiovascular risk assessment. The goals of the NHANES blood lipids measurements include: 1) monitoring the prevalence and trends in major cardiovascular conditions and overall risk factors in the U.S.; 2) evaluating prevention and treatment programs targeting cardiovascular disease in the U.S.; and 3) monitoring the status of hyperlipidemia.
In 2018, new Blood Cholesterol Guidelines were released by the American College of Cardiology and American Heart Association Task Force on Clinical Practice Guidelines, which aim to reduce the risk of atherosclerotic cardiovascular disease through cholesterol management (Grundy, et. al., 2018). The blood lipids measurements in NHANES include total cholesterol, high-density lipoprotein cholesterol (HDL-C), low-density lipoprotein cholesterol (LDL-C), and triglycerides. The present file provides data on HDL-C. Data on total cholesterol are provided in Cholesterol - Total TCHOL_L, and LDL-C and triglyceride data are provided in TRIGLY_L.
Examined participants aged 6 years and older were eligible.
High-Density Lipoprotein (HDL)
This automated HDLC4 assay uses detergents, cholesterol esterase (CHER), cholesterol oxidase (CHOD), and peroxidase to form a colored pigment that is measured optically. The HDLC4 assay meets the 1998 National Institutes of Health (NIH) / National Cholesterol Education Program (NCEP) goals for precision and accuracy.
In this method, non‑HDL lipoproteins such as LDL, VLDL, and chylomicrons are combined with polyanions and a detergent, forming a water‑soluble complex. In this complex, the enzymatic reaction of CHER and CHOD towards non‑HDL lipoproteins is blocked. Finally, only HDL‑particles can react with CHER and CHOD. The concentration of HDL‑cholesterol is determined enzymatically by CHER and CHOD. Cholesterol esters are broken down quantitatively into free cholesterol and fatty acids by CHER.
Refer to the Laboratory Method Files section for a detailed description of the laboratory methods used.
Serum HDL-C levels were calculated from directly measured values of total cholesterol, triglycerides, and LDL-C. Please see below the Data Processing and Editing section for more details. For laboratory methods used for total cholesterol, LDL-C, and triglycerides, please refer to the accompanying documentation: TCHOL_L and TRIGLY_L
There were no changes to the lab method or lab site for this component in the NHANES August 2021 - 2023 cycle. However, the lab equipment used changed from the Cobas 6000 Analyzer to the Cobas 8000 during the cycle.
HDL- Cholesterol Laboratory Procedure Manual (September 2024)
Serum specimens were processed, stored, and shipped to the University of Minnesota, Minneapolis, MN for analysis.
Detailed instructions on specimen collection and processing are discussed in the NHANES Laboratory Procedures Manual (LPM). Vials are stored under appropriate frozen (–30°C) conditions until they are shipped to University of Minnesota for testing.
The NHANES quality assurance and quality control (QA/QC) protocols meet the 1988 Clinical Laboratory Improvement Amendments mandates. Detailed QA/QC instructions are discussed in the NHANES LPM.
Mobile Examination Centers (MECs)
Laboratory team performance is monitored using several techniques. NCHS and contract consultants use a structured competency assessment evaluation during visits to evaluate both the quality of the laboratory work and the QC procedures. Each laboratory staff member is observed for equipment operation, specimen collection and preparation; testing procedures and constructive feedback are given to each staff member. Formal retraining sessions are conducted annually to ensure that required skill levels were maintained.
Analytical Laboratories
NHANES uses several methods to monitor the quality of the analyses performed by the contract laboratories. In the MEC, these methods include performing blind split samples collected during “dry run” sessions. In addition, contract laboratories randomly perform repeat testing on 2% of all specimens.
NCHS developed and distributed a QC protocol for all the contract laboratories, which outlined the use of Westgard rules (Westgard et. al., 1981) when running NHANES specimens. Progress reports containing any problems encountered during shipping or receipt of specimens, summary statistics for each control pool, QC graphs, instrument calibration, reagents, and any special considerations are submitted to NCHS quarterly. The reports are reviewed for trends or shifts in the data. The laboratories are required to explain any identified areas of concern.
The data were reviewed. Incomplete data or improbable values were sent to the performing laboratory for confirmation.
One derived variable was created in this data file. The variable was created using the following formula:
LBDHDDSI
HDL-cholesterol in mg/dL (LBXHDD) was converted to mmol/L (LBDHDDSI) by multiplying by 0.02586.
There are over 800 laboratory tests performed on NHANES participants. However, not all participants provided biospecimens or enough volume for all the tests to be performed. The specimen availability can also vary by age or other population characteristics. Analyst should evaluate the extent of missing data in the dataset related to the outcome of interest as well as any predictor variables used in the analyses to determine whether additional re-weighting for item non-response is necessary.
Please refer to the NHANES Analytic Guidelines and the on-line NHANES Tutorial for further details on the use of sample weights and other analytic issues.
Phlebotomy Weights
For the August 2021-August 2023 cycle, analysis of nonresponse
patterns for the phlebotomy component in the MEC examination revealed
differences by age group and race/ethnicity, among other characteristics. For
example, approximately 67% of children aged 1-17 years who were examined in the
MEC provided a blood specimen through phlebotomy, while 95% of examined adults
aged 18 and older provided a blood specimen. Therefore, an additional
phlebotomy weight, WTPH2YR, has been included in this data release to address
possible nonresponse bias. Participants
who are eligible but did not provide a blood specimen have their phlebotomy weight
assigned a value of “0” in their records. The phlebotomy weight should be used for analyses
that use variables derived from blood analytes, and is included in all relevant
data files.
Demographic and
Other Related Variables
The analysis of NHANES laboratory data must be conducted using the appropriate survey design and demographic variables. The NHANES August 2021-August 2023 Demographics File contains demographic data, health indicators, and other related information collected during household interviews as well as the sample design variables. The recommended procedure for variance estimation requires use of stratum and PSU variables (SDMVSTRA and SDMVPSU, respectively) in the demographic data file.
The Fasting Questionnaire File includes auxiliary information, such as fasting status, length of fast and the time of venipuncture.
This laboratory data file can be linked to the other NHANES data files using the unique survey participant identifier (i.e., SEQN).
Detection Limits
The detection limit was constant for the analyte in the data set.
The lower limit of detection (LLOD, in mg/dL) for HDL-cholesterol:
Variable Name | Analyte Description | LLOD |
---|---|---|
LBXHDD | HDL Cholesterol | 3 |
No Correction Need for HDL-Cholesterol Results for NHANES August 2021 - August 2023
A method validation (bridging) study was performed to compare results from an instrument change in the August 2021-August 2023 cycle. The Cobas 6000 Chemistry Analyzer was upgraded to the Cobas 8000 Chemistry Analyzer during the cycle. Randomly selected serum samples (n=178) from previous NHANES were measured using both instruments and the results were used to conduct the analysis. On average, HDL-Cholesterol values measured from the Cobas 6000 were 0.91% higher than values from the Cobas 8000 (p = 0.0001). Data from the bridging study indicated the correlation coefficient (r) between the measurements was 0.997. Regression analyses were performed using Analyse-it, v4.30.4. The weighted Deming regression equation did not show a significant slope or intercept (95% confidence interval [CI] for slope included 1 and for intercept included 0 (mg/L):
Cobas 8000 = 0.05591 + 0.9894 * (Cobas 6000); 95% CI of slope (0.9780 to 1.001) and intercept (-0.5311 to 0.6429).
Therefore, the NHANES August 2021-August 2023 HDL-Cholesterol values did not have to be adjusted.
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
4391.8220579 to 241728.85724 | Range of Values | 7316 | 7316 | |
0 | No blood sample provided | 752 | 8068 | |
. | Missing | 0 | 8068 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
22 to 159 | Range of Values | 6890 | 6890 | |
. | Missing | 1178 | 8068 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
0.57 to 4.11 | Range of Values | 6890 | 6890 | |
. | Missing | 1178 | 8068 |