Measurement of sodium, potassium, and chloride in a subset of stored casual urine samples from NHANES III (1988-1994).
1,250 participants aged 20-59 years of age selected from a convenience sample of 2,555 casual urine samples collected during NHANES III; selection was based on blood pressure and reported sodium intake levels as described below to ensure a wide range of values for these two related measures.
Samples were selected from a convenience sample of stored urine from NHANES III. To be selected into the sample untreated non-hypertensive respondents had to meet the criteria listed below:
This group of untreated non-hypertensives was then stratified into 8 groups based on their blood pressure category and sodium intake from one dietary recall. Based on these 8 strata sample persons were randomly selected with equal numbers per strata. The untreated hypertensives that fit the above criteria were all included.
Measurement of Urine Sodium, Potassium, and Chloride
All assays were conducted using the Cobas ISE/Na+, K+, Cl- assay performed on the Roche Modular P instrument. The measurement of electrolytes in urine via ion-selective electrode (ISE) is an established and widely used procedure in routine clinical laboratories. The ISE makes use of the unique properties of certain membrane materials to develop an electrical potential for the measurement of ions in solution. The method quantitatively determines sodium, potassium and chloride in urine. The manufacturer has standardized this method against primary calibrators prepared gravimetrically from purified salts. Two levels of standard are provided by the manufacturer for daily calibration. The volume of urine needed per test is 150 µL (combined measurement of sodium, potassium, and chloride).
Analyte name | Analyte code | LOD (mmol/L) |
---|---|---|
Urine sodium (Na+) | UNA | 10 |
Urine potassium (K+) | UK | 1.5 |
Urine chloride (Cl-) | UCL | 10 |
Data were received after all the laboratory testing was complete. The data were not edited for extreme values.
Data will only be available through the Research Data Center.
The laboratory and method were certified according to the Clinical Laboratory Improvement Amendment (1988) guidelines.
For urine Na+, K+ and Cl- , commercially prepared quality control material was purchased from CLINIQA in two levels (LIQUID QCTM 1 and LIQUID QCTM 2). The controls are liquid and ready for use. Stability of the constituents at 2-8°C is 36 months or until the expiration date, whichever comes first. Bench QC pools were characterized to determine the mean and control limits by analyzing duplicate samples in at least 20 runs. After establishment of the control limits of the pools, QC samples contained within each analytical run were evaluated for validity by use of a multi-rule quality control program.
QC Pool | Analyte | N | Mean (mmol/L) | SD (mmol/L) | CV (%) |
---|---|---|---|---|---|
LIQUID QCTM 1 | UNA | 47 | 70.3 | 1.6 | 2.30% |
LIQUID QCTM 2 | UNA | 47 | 163 | 2.11 | 1.30% |
LIQUID QCTM 1 | UK | 47 | 27.8 | 0.87 | 3.10% |
LIQUID QCTM 2 | UK | 47 | 73.8 | 3.37 | 4.60% |
LIQUID QCTM 1 | UCL | 47 | 82.1 | 1.9 | 2.30% |
LIQUID QCTM 2 | UCL | 47 | 193 | 4.02 | 2.10% |
The laboratory achieved satisfactory participation in the CAP proficiency testing surveys during 2010 for the urine chemistry surveys for sodium, potassium, and chloride and for the linearity and calibration verification surveys for sodium.
The laboratory also included blind QC samples (high or low pool) in each run at a rate of 1 blind QC in every 20 samples. These samples were labeled the same as unknown participant samples and therefore remained unrecognized by the analyst.
Blind QC Pool | Analyte | N | Mean (mmol/L) | SD (mmol/L) | CV |
---|---|---|---|---|---|
Low | UNA | 33 | 169 | 1.67 | 1.00% |
High | UNA | 33 | 59.9 | 1.64 | 2.70% |
Low | UK | 33 | 73 | 2.83 | 3.90% |
High | UK | 33 | 36.3 | 0.77 | 2.10% |
Low | UCL | 33 | 187 | 2.5 | 1.30% |
High | UCL | 33 | 57.5 | 2.52 | 4.40% |
This project did not include measurement of urine creatinine because data is already available. However to evaluate the quality of the approximately 20 year old specimens (potential sample evaporation during long-term frozen storage was a concern), we measured creatinine in 300 specimens (selected so that the original concentrations were distributed throughout the sample person range of values) and assessed the correlation and agreement between the new and original results. We obtained reasonable correlation (r = 0.83) and the new results were on average 12% lower than the original results. We therefore concluded that the long-term frozen storage didn’t appear to have led to sample evaporation and the differences seen were probably a result of small method changes over the years.
The subsample of participants aged 20-59 y from NHANES III requires modification of original sampling weights. New sampling weights were generated by NCHS. Weights were created to account for non-response and to sum to the examination weights. The adjustment factors were based on the selection group (1-9) and race/ethnicity. The selection group variable was used since for each of the 9 groups there were different selection probabilities. Given the small cell sizes only one demographic variable could be added to adjust the weights.
Several sensitivity analyses were run to assess the weights. All methods were quite robust. The method chosen had the least variability of the weights.
All statistical analyses using these data should assess the relative standard error of the estimates, the degrees of freedom and the design effect. The three diagnostics listed above are important to assess the reliability of the estimates.
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
7.1 to 358 | Range of Values | 1250 | 1250 | |
. | Missing | 0 | 1250 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
3.2 to 186 | Range of Values | 1250 | 1250 | |
. | Missing | 0 | 1250 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
7.1 to 363 | Range of Values | 1250 | 1250 | |
. | Missing | 0 | 1250 |
Code or Value | Value Description | Count | Cumulative | Skip to Item |
---|---|---|---|---|
4573.0909 to 1054922.6152 | Range of Values | 0 | 0 | |
4573.0909163 | 4573.0909163 | 1250 | 1250 | |
. | Missing | 0 | 1250 |